Doctoral Thesis
DOI
https://doi.org/10.11606/T.5.2004.tde-20082007-075149
Document
Author
Full name
Carlos Augusto Cardoso Pedra
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2004
Supervisor
Committee
Sousa, José Eduardo Moraes Rego (President)
Alves, Claudia Maria Rodrigues
Atik, Edmar
Lucchese, Fernando Antonio
Souza, José Augusto Marcondes de
Title in Portuguese
Análise dos resultados imediatos e tardios do tratamento percutâneo da coartação da aorta em adolescentes e adultos: comparação entre balões e stents
Keywords in Portuguese
adolescência
adulto
angioplastia com balão/efeitos adversos
angioplastia com balão/métodos
avaliações de resultado de intervenções terapêuticas
cateterismo/efeitos adversos
Cateterismo/métodos
Coartação aórtica
Coartação aórtica/terapia
Contenedores/efeitos adversos
Grupos controle
seguimentos
Abstract in Portuguese
Mais informações são necessárias para definir se o tratamento percutâneo da coartação da aorta com stents é superior à angioplastia com cateter-balão. De julho de 2000 a maio de 2003, 21 adolescentes e adultos com coartação da aorta focal e média de idade de 24 anos (DP 11 anos) foram submetidos, consecutivamente, a implante de stents (grupo 1). Os resultados foram comparados com os obtidos em um grupo histórico de 15 pacientes com média de idade de 18 anos (DP 10 anos) (p = 0,103) submetidos a angioplastia (grupo 2) nos últimos 18 anos. Após o procedimento, a redução do gradiente sistólico foi maior (99% [DP 2%] versus 87% [DP 17%]; p = 0,015), o gradiente residual foi menor (0,4 mmHg [DP 1,4 mmHg] versus 5,9 mmHg [DP 7,9 mmHg]; p = 0,019), o ganho no local da coartação foi maior (333% [DP 172%] "versus" 190% [DP 104%]; p = 0,007) e o diâmetro da coartação foi maior (16,9 mm [DP 2,9 mm] versus 12,9 mm [DP 3,2 mm]; p < 0,001) no grupo 1. Alterações da parede da aorta, incluindo dissecções, abaulamentos e aneurismas, foram observadas em oito pacientes do grupo 2 (53%) e em um do grupo 1 (7%) (p < 0,001). Não houve complicações maiores. Cateterismo (n = 33) ou ressonância magnética (n = 2) de controle foram realizados em seguimento mediano de um ano para o grupo 1 e um ano e meio para o grupo 2 (p = 0,005). A redução do gradiente sistólico persistiu em ambos os grupos; entretanto, gradientes tardios mais altos foram observados no grupo 2 (mediana de 0 mmHg para o grupo 1 versus 3 mmHg para o grupo 2; p = 0,014). Não houve perdas no diâmetro da coartação no grupo 1 e houve ganho tardio no grupo 2 (16,7 mm [DP 2,9 mm] versus 14,6 mm [DP 3,9 mm]; p = 0,075). No grupo 1, dois pacientes necessitaram de novo implante de stent em decorrência da formação de aneurisma ou fratura da malha do stent. No grupo 2, as anormalidades da parede aórtica não progrediram e um paciente necessitou de redilatação em decorrência da recoartação. A pressão arterial sistêmica foi semelhante em ambos os grupos durante o seguimento (sistólica: 126 mmHg [DP 12 mmHg] no grupo 1 versus 120 mmHg [DP 15 mmHg] no grupo 2; diastólica: 81 mmHg [DP 11 mmHg] no grupo 1 versus 80 mmHg [DP 10 mmHg] no grupo 2; p = 0,149 e p = 0,975, respectivamente). Apesar de os desfechos clínicos terem sido satisfatórios e similares com ambas as técnicas, o uso de stents propiciou resultados mais previsíveis e uniformes para alívio da estenose, minimizando também o risco de desenvolvimento de alterações da parede da aorta.
Title in English
Analysis of the immediate and late results of percutaneous treatment of coarctation of the aorta in adolescents and adults: comparison between balloons and stents
Keywords in English
Adolescence
Adult
Angioplasty baloon/adverse effects
Angioplasty baloon/methods
Aortic coarctation
Aortic coarctation/therapy
Catheterization/methods
Control groups
Evaluation of results of therapeutic interventions
Follow-up studies
Stents/adverse effects
Abstract in English
More information is needed to define whether stenting is superior to balloon angioplasty for coarctation of the aorta. From July/2000 to May/2003, 21 adolescents and adults with discrete coarctation underwent consecutive stent implantation at a mean age of 24 years (SD 11 years) (group 1). The results were compared to those achieved by balloon angioplasty performed in the last 18 years in a historical group of 15 patients at a mean age of 18 years (SD 10 years) (p = 0.103) (group 2). After the procedure, systolic gradient reduction was higher (99% [SD 2%] vs. 87% [SD 17%]; p = 0.015), residual gradients lower (0.4 mmHg [SD 1.4 mmHg] vs. 5.9 mmHg [SD 7.9 mmHg); p = 0.019), gain at the coarctation site higher (333% [SD 172%] vs. 190% [SD 104%]; p = 0.007) and coarctation diameter larger (16.9 mm [SD 2.9 mm] vs.12.9 mm [SD 3.2 mm]; p < 0.001) in group 1. Aortic wall abnormalities, including dissections, bulges and aneurysms, were observed in eight patients in group 2 (53%) and in one in group 1 (7%) (p < 0.001). There was no major complication. Repeat catheterization (n = 33) or magnetic resonance imaging (n = 2) was performed at a median follow-up of 1.0 year for group 1 and 1.5 year for group 2 (p = 0.005). Gradient reduction persisted in both groups, although higher late gradients were observed in group 2 (median of 0 mmHg for group 1 vs. 3 mmHg for group 2; p = 0.014). There was no late loss in the coarctation diameter in group 1 and there was a late gain in group 2 (16.7 mm [SD 2.9 mm] for group 1 vs. 14.6 mm [SD 3.9 mm] for group 2; p = 0.075). Two patients required late stenting due to aneurysm formation or stent fracture in group 1. Aortic wall abnormalities did not progress and one patient required redilation due to recoarctation in group 2. Blood pressure was similar in both groups at follow-up (systolic: 126 mmHg [SD 12 mmHg] in group 1 vs. 120 mmHg [SD 15 mmHg] in group 2; diastolic: 81 mmHg [SD 11 mmHg] in group 1 vs. 80 mmHg [SD 10 mmHg] in group 2; p = 0.149 and p = 0.975, respectively). Although satisfactory and similar clinical outcomes were observed with both techniques, the use of stents yielded more predictable and uniform results for stenosis relief, also minimizing the risk of developing aortic wall abnormalities.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
stentCoAo.pdf (4.84 Mbytes)
Publishing Date
2007-09-17
 