Thèse de Doctorat
DOI
https://doi.org/10.11606/T.46.2011.tde-18082011-141833
Document
Auteur
Nom complet
Juan Claudio Mancilla Gamboa
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2011
Directeur
Jury
Bertotti, Mauro (Président)
Masini, Jorge Cesar
Muñoz, Rodrigo Alejandro Abarza
Santos, Mauro Coelho dos
Serrano, Silvia Helena Pires
Titre en portugais
Fabricação de sensores eletroquímicos para a determinação de espécies químicas de relevância ambiental
Mots-clés en portugais
Amônio
Eletrodo de cobre
Eletrodo de íon seletivo
Nitrato
Nitrito
Resumé en portugais
Neste trabalho são apresentados resultados relacionados à determinação de nitrato, nitrito e amônio por técnicas eletroquímicas. Para a determinação de nitrato foi utilizada uma lâmina de cobre cuja superfície é renovada continuamente empregando um processo de ativação (oxidação do cobre e subseqüente redução de íons Cu(II)) com potencial controlado em sistema FIA. As condições ótimas foram alcançadas por meio de 60 s para a etapa de ativação, vazão do transportador 3,0 mL min-1 e volume de amostra 150 µL. A repetibilidade foi estimada em 4,7% (n=9) com freqüência de amostragem de 60 amostras h-1. Trabalhando nas condições otimizadas um aumento na concentração de nitrato gerou um aumento proporcional na corrente, resultando em curva analítica em um intervalo de 0,1 2,5 mmol L-1 (-Ip (µA) = 0,13 + 4,6 [NO3-] (mmol L-1), R2 = 0,9991). Os limites de detecção e quantificação foram estimados em 4,2 (S/N = 3) e 14 (S/N = 10) µmol L-1, respectivamente. O mesmo sensor foi utilizado para determinações de nitrito por voltametria de pulso diferencial e a otimização dos parâmetros relativos ao processo de ativação (tempos de dissolução e redeposição) foi efetuada com o intuito de aumentar a sensibilidade. As condições ótimas foram de 30 s para a etapa de ativação, com uma repetibilidade de 1,9% (n=10), e o aumento nas concentrações de nitrito gerou um aumento proporcional na corrente, resultando em curva analítica em um intervalo de 50 µmol L-1 - 1,44 mmol L-1 (-Ip (µA) = -0,13 + 53,52 [NO2-] (mmol L-1 ), R2 = 0,9996). Os limites de detecção e quantificação foram estimados em 2,8 (S/N = 3) e 9,4 (S/N = 10) µmol L-1, respectivamente. O estudo da morfologia da superfície do eletrodo de cobre durante o processo de ativação foi realizado por AFM com o objetivo de compreender o impacto das mudanças microestruturais no sinal de corrente para nitrato e nitrito. Os resultados indicam que após a ativação do eletrodo a superfície apresenta cristais com uma textura rugosa e ocorre aumento da área superficial, justificando o aumento de sinal de corrente. Para a determinação de amônio foram fabricados eletrodos íon seletivo (ISE) com membrana polimérica de forma tubular, os quais foram acoplados em um sistema de injeção em fluxo. A influência de parâmetros como vazão do transportador e alça de amostragem foi investigada e melhores resultados foram obtidos em 0,5 mL min-1 e 250 µL, respectivamente. Nas condições otimizadas a repetibilidade das determinações foi de 1,3 % (n=10) com freqüência de amostragem de 13 amostras h-1. Tendo em conta as otimizações do sistema FIA, injeções sucessivas de soluções de amônio com concentrações crescentes permitiram obter uma reta com coeficiente angular de 51,2 mV (R2 = 0,9836) no intervalo de 0,2 5 mg L-1. Os limites de detecção foram calculados em 0,05 mg L-1, respectivamente.
Titre en anglais
Fabrication of electrochemical sensors for determination of chemical species of environmental relevance
Mots-clés en anglais
Ammonium
Copper of electrode
Ion selective electrode
Nitrate
nitrite
Resumé en anglais
This paper presents results related to the determination of nitrate, nitrite and ammonium by electrochemical techniques. A copper electrode whose surface is continuously renewed using an activation process with controlled potential in a FIA system was used for nitrate determinations. The optimum conditions were achieved by an activation step of 60 s, carrier stream of 3.0 mL min-1 and sample volume of 150 µL. The repeatability was estimated as 4.7% (n = 9) with a sampling frequency of 60 samples h-1. At these optimized experimental conditions an increase in nitrate concentration caused a proportional increase in current, resulting in an analytical curve in the range from 0.1 to 2.5 mmol L-1 (-Ip (µA) = 0.13 + 4.6 [NO3-] (mmol L-1), R2 = 0.9991). The limits of detection and quantification were estimated as 4.2 (S/N = 3) and 14 (S/N = 10) µmol L-1, respectively. The same sensor was used for the determination of nitrite by differential pulse voltammetry and the optimization of the parameters involved in the activation process (dissolution and redeposition time) was performed in order to increase the sensitivity. Optimum conditions were established at 30 s for the activation step, with a repeatability of 1.9% (n = 10) and increased concentrations of nitrite generated a proportional increase in current, resulting in an analytical curve in the range from 50 µmol L-1 to 1.44 mmol L-1 (-Ip (µA) = -0.13 + 53.52 [NO2-] (mmol L-1), R2 = 0.9996). The limits of detection and quantification were estimated as 2.8 (S/N = 3) and 9.4 (S/N = 10) µmol L-1, respectively. The study of the morphology of the copper electrode surface upon the activation process was performed by AFM in order to understand the influence of microstructural changes in the current signal for nitrate and nitrite. The results indicated that after the activation the electrode surface presented a rough texture with a concurrent increase in the surface area, justifying the current increase. Ion-selective eletrodes (ISE) with a polymeric membrane in a tubular shape were fabricated for the determination of ammonium in a flow injection system. The influence of parameters such as the carrier stream and the sample volume was investigated and best results were obtained at 0.5 mL min-1 and 250 mL, respectively. Under optimum conditions the repeatability of the determinations was 1.3% (n = 10) with a sampling frequency of 13 samples h-1. Taking into account the optimization of the FIA system, successive injections of solutions with increasing concentrations of ammonium yielded a straight line with slope of 51.2 mV (R2 = 0.9836) in the range 0.2 to 5 mg L-1. The limits of detection were calculated as 0.05 mg L-1, respectively
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2011-08-30
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.