Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.46.2012.tde-10092012-095528
Document
Auteur
Nom complet
Valeska Meirelles Sousa
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2012
Directeur
Jury
Oliveira, Pedro Vitoriano de (Président)
Gaubeur, Ivanise
Paixão, Thiago Regis Longo Cesar da
Titre en portugais
Desenvolvimento de método para a determinação de irídio, paládio, platina e ródio por ICP OES em sílica mesoporosa para uso catalítico
Mots-clés en portugais
Catalisador automotivo
Espectrometria de emissão óptica com plasma indutivamente acoplado
Padrão interno
Paládio
Platina
Preparo de amostra
Ródio
Resumé en portugais
O objetivo deste estudo foi avaliar parâmetros tais como, moagem, tamanho de partículas, homogeneidade, segregação, dissolução de amostras e amostragem de suspensão, visando o desenvolvimento de método para a determinação de Ir, Pd, Pt e Rh por espectrometria de emissão óptica com plasma indutivamente acoplado (ICP OES) em sílica mesoporosa para uso catalítico. Após a moagem em moinho eletrônico de almofariz e pistilo, a análise do tamanho e distribuição das partículas revelou que 89% das partículas apresentam tamanhos <100 µm e 23% com tamanho <5 µm. Na avaliação de segregação dos analitos na amostra, os resultados obtidos para amostra com moagem grosseira, indicaram que os elementos estão distribuídos de maneira heterogênea entre as frações. Entretanto, a moagem fina aumentou a homogeneidade e esta pode ser uma etapa importante e necessária no sentido de evitar segregação da amostra. A análise por difração e de Raios-X mostrou que não ocorreram reações mecanoquímicas durante a moagem e por fluorescência de Raios-X forneceu resultados discrepantes entre os laboratórios, o pode estar relacionado com a heterogeneidade que induziu erros de amostragem. A avaliação de desempenho do ICP OES foi feita a partir do estabelecimento de parâmetros de mérito como seletividade, repetibilidade, estabilidade, robutez e limite de detecção. Na otimização foram considerados os parâmetros potência, vazão de nebulização e vazão de gás auxiliar. Foi feito um estudo em solução aquosa e em meio da amostra para escolha das linhas de emissão dos elementos Ir, Pd, Pt, Rh e dos candidatos a padrões internos. Apesar dos bons resultados encontrados em solução aquosa, o aumento do sinal de fundo em meio da amostra restringiu a escolha dos comprimentos de onda. Para o estudo da dissolução das amostras foram avaliados quatro procedimentos em forno de micro-ondas com frasco fechado. Os melhores resultados foram obtidos com 150 mg de amostra, 6 mL de água régia e 210 ºC de aquecimento durante 60 min. As determinações por ICP OES usando padrões internos Ru para Pd, Yb para Pt e Dy para Rh foram aquelas que forneceram os resultados mais coerentes com aqueles obtidos por fluorescência de Raios-X. Nos estudos visando à introdução de amostra na forma de suspensão foram avaliados: a massa de amostra para o preparo da suspensão, o efeito da concentração de ácido nítrico e o uso de padrão interno na tentativa de corrigir erros de matriz. 60 mg de amostra com um volume final de 20 mL de ácido nítrico 0,5% (v v-1) foram os escolhidos. Entretanto, a determinação dos elementos via suspensão mostrou resultados muito inferiores aos valores de concentrações esperados. Irídio foi investigado somente para otimização de ICP OES porque a sílica mesoporosa para uso de catalisador automotivo tem não esse elemento em sua composição.
Titre en anglais
Development of method for the determination of iridium, palladium, platinum and rhodium in mesoporosa silica for catalyst use by ICP OES
Mots-clés en anglais
Catalyst
Inductively coupled plasma optical emission spectrometry
Internal standard
Iridium
Platinum
Rhodium
Sample preparation
Resumé en anglais
The aim of this study was to evaluate parameters such as grinding, particle size, homogeneity, segregation, sample dissolution and slurry sampling for method development for the determination of Ir, Pd, Pt and Rh in silica mesoporosa for catalyst using inductively coupled plasma optical emission spectrometry (ICP OES). After the grind in an electronic grider with mill and pistil, the analysis of the particle size distribution showed 89% of particles with sizes <100 µm and 23% of particles with sizes <5µm. The separation in different particle size fractions showed that the sample with coarse grind (as sent to the lab) is not homogenous for the elements in comparison with that submitted to grind. X-Ray Diffraction and X-Ray Fluorescence analysis did not showed chemical reactions during the grinding, however a discrepancy between results of different laboratories. The performance of ICP OES was made by using figures of merit such as selectivity, repeatability, stability, robustness and limit of detection. In the optimization of ICP OES power supply, nebulizer flow rate and auxiliary gas flow rate were considered. Studies were conducted in aqueous and in the sample solution to choose emission wavelengths for Ir, Pd, Pt, Rh and internal standard candidates. Despite the good results obtained in aqueous solution, increasing the background signal in sample media restricted the choice of the emission wavelengths. For sample dissolution four procedures using microwave oven with closed-vessel were evaluated. The best results were obtained with 150 mg of the sample, 6 mL of aqua-regia and 210 ºC of heating, during 60 min. Determinations by ICP OES using internal standards Ru for Pd, Yb for Pt, and Dy for Rh were those who provided more consistent results in comparison with those obtained by x-ray fluorescence. For slurry sampling introduction sush parameters were evaluated: the mass of sample to prepare the slurry, the effect of the nitric acid concentration and the use of internal standard to correct matrix effects. 60 mg of the sample with a final volume of 20 mL of nitric acid 0.5% (v v-1) were chosen. However, the determination of Pd, Pt and Rh via suspension showed concentrations below to those expected, as determined by ICP OES in solution. Iridium was investigated only for ICP OES optimization because the silica mesoporosa for automobile catalyst use have not this element in its composition.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2012-10-29
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.