Master's Dissertation
DOI
https://doi.org/10.11606/D.46.2018.tde-21092018-095836
Document
Author
Full name
Alexandra Iwazaki
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2003
Supervisor
Committee
Tokoro, Roberto (President)
Bertotti, Mauro
Pezza, Leonardo
Title in Portuguese
Estudo da influência da força iônica na medição de H+ utilizando eletrodo de membrana de vidro
Keywords in Portuguese
Calibração do eletrodo
Concentração hidrogenionica
Força iônica
Membrana de vidro
Método eletroquímico (Aplicações)
Ph
Potenciometria (Aplicações; Estudo)
Química analítica instrumental
Abstract in Portuguese
O presente trabalho mostra a influência da força iônica na leitura potenciométrica de pH de soluções aquosas utilizando membrana de vidro. O tamanho e a carga dos íons são importantes fatos apontados. Um tipo de efeito de "memória" aparece durante as leituras de pH, este foi contornado pela calibração multipontuada. Propostas para corrigir a influência da força iônica foram estudadas sob aspectos experimental e teórico. A teoria de Cheng sobre o mecanismo de ação da membrana de vidro como capacitor químico foi comprovada por estudos em que se variou a ação do campo elétrico. O conceito da grandeza de pH também foi discutido.
Title in English
Study of the influence of the ionic force on the measurement of H + using glass membrane electrode
Keywords in English
Electrochemical method (Applications)
Electrode calibration
Glass membrane
Hydrogenion concentration
Instrumental analytical chemistry
Ionic force
Ph
Potentiometry (Applications
Study)
Abstract in English
This work shows the influence of the ionic strength on the potenciometric measurements of pH of aqueous solutions using the electrode of glass membrane. The size and charge of the ions are important studied facts. One type of "memory" of glass membrane was observed and this effect was overcame using multipoint calibration. To correct the influence of the ionic strength two studies on experimental and theorical aspects were performed. The model proposed by Cheng to the mechanism of glass membrane as a chemical capacitor was confirmed by application of electrical field on the glass membrane. The concept of pH amount was discussed.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2018-09-21
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.