Thèse de Doctorat
DOI
https://doi.org/10.11606/T.46.2001.tde-17102002-122859
Document
Auteur
Nom complet
José Alberto Fracassi da Silva
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2001
Directeur
Jury
Lago, Claudimir Lucio do (Président)
Carrilho, Emanuel
Gutz, Ivano Gebhardt Rolf
Pasquini, Celio
Tavares, Marina Franco Maggi
Titre en portugais
Detecção condutométrica sem contato (oscilométrica) para eletroforese capilar de zona e cromatografia micelar eletrocinética
Mots-clés en portugais
detecção eletroquímica
eletroforese capilar
instrumentação
Resumé en portugais
Este trabalho descreve a construção e avaliação de um detector condutométrico sem contato (oscilométrico) para sua aplicação em eletroforese capilar de zona e cromatografia micelar eletrocinética (MEKC). A construção do detector contou com a avaliação de diversos materiais e métodos para a confecção dos eletrodos, tão bem como o aperfeiçoamento do seu circuito eletrônico. O seu comportamento e desempenho foram verificados através do estudo dos diversos parâmetros que influenciam sua resposta, como freqüência e amplitude do sinal aplicado, temperatura e condutividade do meio. Além disso, a simulação do circuito equivalente da cela de detecção auxiliou no entendimento das propriedades do detector frente a alterações na condutividade do meio, na freqüência de operação e nas dimensões da cela. A otimização dos parâmetros operacionais foi racionalizada pela formulação de equações analíticas que descrevem o fator de resposta do detector a partir de parâmetros obtidos experimentalmente. Para o desenvolvimento do sistema de detecção, dois equipamentos completos de eletroforese capilar foram construídos. Sistemas de injeção de amostra por pressão, por gravidade, e eletrocinética foram desenvolvidos. Um dos equipamentos permite que a injeção da amostra seja feita do lado aterrado da fonte de alta tensão. Conseqüentemente, o detector é posicionado próximo do ponto de aplicação da alta tensão. Este é, de que se tem notícia, o primeiro sistema de detecção eletroquímico para eletroforese capilar com essa característica. A termostatização dos capilares foi efetuada por convecção forçada de ar. De maneira a aumentar o grau de automação e facilitar o seu manuseio, os dois equipamentos foram interfaceados a microcomputador. Os equipamentos de eletroforese com detecção oscilométrica construídos permitiram a análise de diversas substâncias em matrizes variadas, como por exemplo íons orgânicos e inorgânicos em água de chuva e material particulado, ácidos graxos de cadeia linear em gordura de coco, formaldeído na fase gasosa da atmosfera, e, pela primeira vez, espécies neutras, como álcoois alifáticos, separadas por MEKC.
Titre en anglais
Contactless conductivity detection for capillary zone electrophoresis and micellar electrokinetic chromatography
Mots-clés en anglais
capillary electrophoresis
electrochemical detection
instrumentation
Resumé en anglais
This work describes the construction and evaluation of a contactless conductometric (or oscillometric) detector, for its application in capillary zone electrophoresis and micellar electrokinetic chromatography (MEKC). The materials and methods used for the construction of the electrodes, as well as the improvements on the electronic circuit, were evaluated. The behavior and performance of the detector were studied by changing the parameters that affect its response, e. g., the frequency and amplitude of the input signal, temperature and solution conductivity. The simulation of the cell equivalent electric circuit helped to understand the detector properties related to the variations in the solution conductivity, frequency, and cell dimensions. The optimization of the parameters was conducted by the formulation of analytical equations that describe the response factor using experimental data. For the detection development, two complete capillary electrophoresis equipments were constructed. Pressure, gravity, and electrokinetic sample injection systems were developed. One of the equipments allows the sample injection on the grounded side of the capillary. Consequently, the detector is placed near the high voltage application point. This is the first electrochemical detector described for capillary electrophoresis with this characteristic. The capillary thermostating was made by passing a forced air stream. Both equipments were interfaced to microcomputers. The capillary electrophoresis equipments with oscillometric detection were applied to the analysis of many types of compounds in different matrices, such as organic and inorganic ions in rainwater and particulate material, fatty acids in cocoa oil, formaldehyde in the atmosphere and, for the first time, neutral species, like aliphatic alcohols, separated by MEKC.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Fracassi.pdf (4.85 Mbytes)
Date de Publication
2003-08-14
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.