Disertación de Maestría
DOI
https://doi.org/10.11606/D.46.2019.tde-15022019-113640
Documento
Autor
Nombre completo
Ruben Gregorio Moreno Moreno
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2001
Director
Tribunal
Oliveira, Elisabeth de (Presidente)
Cadore, Solange
Gutz, Ivano Gebhardt Rolf
Título en portugués
Determinação de mercúrio em amostras ambientais por espectrometria de absorção atômica com atomização eletrotérmica em forno de grafite com superfície modificada
Palabras clave en portugués
Atomic absorption spectrometry
CVAAS
Paládio
Permanent modifier
Química ambiental
Química analítica instrumental
Química analítica qualitativa
Resumen en portugués
Um procedimento eletroquímico para deposição de paládio no interior de um tubo de grafite para modificação química permanente e um sistema de geração de vapor frio para a determinação de níveis traços de mercúrio em água e sedimento através de espectrometria de absorção atômica é proposto. A célula eletroquímica tubular para as deposições de paládio em fluxo foi construída aproveitando a estrutura original do tubo de grafite o qual opera como eletrodo de trabalho. Um tubo de aço inoxidável, posicionado na saída da célula, é usado como o eletrodo auxiliar. O valor do potencial aplicado no eletrodo de grafite é medido contra um micro eletrodo de referência de Ag/AgCl inserido sobre o eletrodo auxiliar. Soluções de paládio em tampão de acetato (100 milimol L-1, pH=4,70), numa vazão de 0,5 mL min-1 foram usados para executar a deposição eletroquímica durante o intervalo de tempo de 60 min. O sistema de geração de vapor frio consiste de um micro-reator de polietileno e frasco separador líquido-gás, volume total de 4,0 ml, que é descartado depois de cada amostragem para evitar inter-contaminação das amostras. Volumes de 1,0 ml de reagente (2,0% m/v NaBH4 em 0,10 mol L-1 de NaOH) e 1,0 ml de soluções analíticas ou amostras em 0,25 mol L-1 de HNO3 são levados para o reator e separador líquido-gás usando uma bomba peristáltica. O vapor de mercúrio é transportado ao tubo de grafite modificado eletroquimicamente com fluxo de argônio de 212 ml min-1 e pré-concentrado durante 120 s. O limite de detecção obtido foi 93 ng L-1 (n=20, 3δ). O desempenho deste sistema foi testado para determinação de mercúrio em água potável, água não potável e sedimentos de lagoa. A confiança de todo procedimento foi confirmada através de testes de recuperação.
Título en inglés
Determination of mercury in environmental samples by atomic absorption spectrometry with electrothermal atomization in graphite furnace with modified surface
Palabras clave en inglés
Atomic absorption spectrometry
CVAAS
Environmental chemistry
Instrumental analytical chemistry
Palladium
Permanent modifier
Qualitative analytical chemistry
Resumen en inglés
An electrochemical procedure for palladium deposition on the inner of a graphite tube for permanent chemical modification and a cold vapor generation system for the determination of trace levels of mercury in water and sediment by atomic absorption spectrometry are proposed. The tubular electrochemical cell for the deposition of palladium was assembled on the original geometry of a graphite tube that operates as the working electrode. A stainless steel tube, positioned downstream from the working electrode, is used as the auxiliary electrode. The potential value applied on the graphite electrode is measured against a micro reference electrode (Ag/AgCl) inserted into the auxiliary electrode. Palladium solutions in acetate buffer (100 milimol L-1, pH=4.70), flowing at 0.5 mL min-1 for 60 min was used to perform the electrochemical deposition. A cold vapor generation system consist of a micro polyethylene reactor and gas-liquid separator flask, total volume of 4.0 mL, that is discarded after each sampling to avoid intercontamination of the samples. Volumes of 1.0 mL of reagent (2.0% m/v NaBH4 in 0.10 mol L-1 of NaOH) and 1.0 mL of reference or sample solutions in 0.25 mol L-1 of HNO3 are carried out to the reactor and gas-liquid separator by using a peristaltic pump. The mercury vapor is transported to the graphite tube electrochemically modified with argon flow (200 mL min-1) and pre-concentrated during 120 s. The detection limit obtained was 93 ng L-1 (n=20, 3δ). The performance of these system was tested for determination of mercury in potable and non-potable water and lake sediments. The reliability of the entire procedure was confirmed by recovery tests.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2019-02-15
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.