Dissertação de Mestrado
DOI
https://doi.org/10.11606/D.43.2006.tde-28032007-171124
Documento
Autor
Nome completo
Diogo Bernardes Tridapalli
E-mail
Unidade da USP
Área do Conhecimento
Data de Defesa
Imprenta
São Paulo, 2006
Orientador
Banca examinadora
Pascholati, Paulo Reginaldo (Presidente)
Added, Nemitala
Koskinas, Marina Fallone
Título em português
Estudo da função resposta de um detetor HPGe a fótons da reação 19F(p,alfa gama)16O
Palavras-chave em português
Fótons de alta energia
Função resposta
HPGe
Monte Carlo
Resumo em português
O Laboratório do Acelerador Linear do Instituto de Física da Universidade de São Paulo está desenvolvendo um estudo sobre a eficiência e a função resposta de detetores de HPGe a fótons de alta energia. Neste trabalho foi estudada a função resposta de fótons provenientes da reação 19F(p,alfa gama)16O. Na reação podem ser observados os fótons característicos de 6,1 MeV, 6,9 MeV e 7,1 MeV. Fótons com esta energia possuem um grande potencial para várias aplicações importantes, como "Proton Induced Gamma-ray Emission" (PIGE), gamagrafia e procedimentos de calibração. No caso de procedimentos de calibração essa reação possui um diferencial: o fluxo de fótons observado é bem maior do que em outras reações estudadas, tais como 27Al (p, gama)28Si e 23Na(p, gama)24Mg. Os experimentos foram realizados no Laboratório de Análise de Materiais por Feixes Iônicos (LAMFI) do IFUSP, utilizando um acelerador eletrostático tipo "Pelletron-tandem" de 1,7 MV de tensão máxima no terminal. O detetor estudado foi um HPGe do tipo coaxial "reverse-electrode closed-end" com 72,5 mm de diâmetro e 60,5 mm de comprimento e 60% da eficiência de um detetor de NaI para fótons de 1,332 MeV. O detetor foi posicionado a 0 grau em relação à linha de feixe. A corrente de prótons na irradiação era da ordem de 50 nA e a energia dos prótons variou entre 1,36 MeV e 1,42 MeV. O alvo utilizado consistia em 250 micro g/cm^2 de CaF2 evaporado em um suporte de Ta. Os picos dos três fótons da reação 19F(p,alfa gama)16O apresentaram deformações devido à emissão em vôo dos fótons. Essas deformações apresentam variações em função da energia dos prótons. Com o objetivo de descrever as formas dos picos foi desenvolvido um programa que utiliza a técnica de Monte Carlo para simulá-las. Os dados sobre "yields" relativos dos fótons e seção de choque da reação 19F(p,alfa gama)16O encontrados na literatura são imprecisos e contraditórios. Então, como parte dos resultados deste trabalho, foram obtidos os "yields" relativos dos fótons da reação 19F(p,alfa gama)16O para energias de prótons entre 1,360 MeV e 1,420 MeV. A função resposta do detetor para as diferentes formas dos picos foram obtidas através de simulações com o programa MCNP5. Para estas simulações foi feita uma descrição detalhada da geometria interna do detetor, obtida por meio de radiografias do mesmo. As intensidades relativas dos fótons da reação 19F(p,alfa gama)16O em função da energia dos prótons puderam ser obtidas através de um ajuste por mínimos quadrados das funções respostas simuladas aos espectros experimentais de altura de pulso.
Título em inglês
Study of HPGe detector response function for photons from the 19F(p, alpha gamma)16O reaction
Palavras-chave em inglês
High energy photons
HPGe
Monte Carlo
Response function
Resumo em inglês
In the 19F(p, alpha gamma)16O reaction, characteristic gamma-rays with energies: 6.1 MeV; 6.9 MeV and 7.1 keV can be observed. These photons can be used in many important applications such as Proton Induced Gamma-ray Emission (PIGE), gamma radiography and calibration purposes. It has another advantage in calibration procedures, which is the observed photon flux is greater than other reactions studied, 27Al(p, gamma)28Si and 23Na(p, gamma)24Mg. In our laboratory, we are studying the efficiency and response function of HPGe detectors for high energy photons, and for this it is necessary a source with a level scheme with a few gamma-ray transitions and known relative yields. The 19F(p, alpha gamma)16O reaction satisfies the first condition but in the literature we found relative yields for thick targets or reaction cross section for thin targets. However, we use targets of intermediate thickness, therefore in this work we measure the relative gamma-ray yields for protons with energies between 1.36 MeV and 1.42 MeV. The experiments were performed at the 1.7 MV Pelletron tandem accelerator of the Laboratório de Análise de Materiais por Feixes Iônicos (LAMFI) located at Instituto de Física da Universidade de São Paulo, using a reverse-electrode closed-end coaxial HPGe detector with 72.5 mm in diameter and 60.5 mm in length, at 0 degree. The proton irradiation current was 50 nA, and the target consisted of 250 micro g/cm^2 CaF2 evaporated on a 0.1 mm Ta backing. In this reaction the 20Ne resonant state fissions in flight, and the resulting 16O excited states have different half-lives; all observed peaks are deformed due to either Doppler broadening from 16O random velocity direction or Doppler shift from nuclei that leave the target toward the vacuum chamber. Moreover, the peak deformations vary with the proton energy. These deformations were modeled by a Monte Carlo simulation that follows the oxygen nuclei in its trajectory until photon emission, considering the changes in spatial distribution of the exact interaction point in the target with incident proton energy due to the large resonance width and proton energy loss, and 16O energy loss and multiple scattering until decay, in the different target layers. Using the detector response functions calculated by MCNP5 simulations, the relative intensities of the three gamma rays were obtained by a least square fit of the response functions, taking into account the Doppler broadening and shift for each gamma ray, to the data in the experimental pulse-height spectrum.
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
DissertacaoDiogoT.pdf (4.33 Mbytes)
Data de Publicação
2007-04-09
 
AVISO: Saiba o que são os trabalhos decorrentes clicando aqui.
Todos os direitos da tese/dissertação são de seus autores
CeTI-SC/STI
Biblioteca Digital de Teses e Dissertações da USP. Copyright © 2001-2024. Todos os direitos reservados.