Dissertação de Mestrado
DOI
https://doi.org/10.11606/D.43.2008.tde-26112008-094456
Documento
Autor
Nome completo
Arlene Linke
E-mail
Unidade da USP
Área do Conhecimento
Data de Defesa
Imprenta
São Paulo, 2008
Orientador
Banca examinadora
Yoshimura, Elisabeth Mateus (Presidente)
Moralles, Mauricio
Tobias, Carmen Cecília Bueno
Título em português
Espectroscopia de raios X utilizando o espalhamento Compton
Palavras-chave em português
Física
Física experiemental
Física médica
Radiologia
Resumo em português
O estudo do espectro de raios X é importante para predizer a qualidade da imagem em sistemas radiográficos. Nesse trabalho usamos um detector semicondutor CdTe para medir os espectros de raios X na faixa de 25 a 120 kVp. Esse detector é pequeno e fácil de ser transportado. As medidas feitas diretamente do tubo de raios X não são aconselháveis devido à alta taxa de fluência que estão acima dos limites de detecção dos detectores usuais e podem causar armadilhamento de carga no detector CdTe. A alternativa é medir os fótons espalhados a 900 a partir de um pequeno objeto espalhador, no nosso caso um cilindro de lucite (PMMA). A partir do espectro espalhado, corrigido para tempo morto e eficiência, e calibrado em energia, efetuamos correções para o ângulo de espalhamento e espalhamento coerente (somente para baixas energias). Só então formalismo de Klein-Nishina para as seções de choque do efeito Compton é aplicado aos espectros, resultando nos espectros primários. Para energia de 120 kVp foi medido um espectro do feixe direto e comparado com o obtido através do espectro espalhado para mesma energia. Para o mesmo feixe, a camada semi-redutora foi avaliada experimentalmente e com o espectro obtido por espalhamento Compton, obtendo-se boa concordância. Observamos na curva de eficiência uma descontinuidade, prevista teoricamente e também observada por outros autores, que deforma o espectro corrigido. Foi aplicado um ajuste polinomial a essa curva, suavizando os espectros obtidos e não alterando sua forma. Os resultados foram satisfatórios e validaram o formalismo apresentado por outros autores, utilizando detectores distintos.
Título em inglês
X-ray spectroscopy using compton scattering
Palavras-chave em inglês
Experimental physics
Medical physics
Physics
Radiology
Resumo em inglês
The analysis of x-ray spectra is important to predict image quality in radiographic systems. In this work we used a semiconductor CdTe detector to measure x-ray spectra with a CdTe detector, in the range from 25 to 120 kVp. The CdTe detector is small and portable. The measurements done directly in the primary X rays beam can be difficult because of the high photon fluence rates that cause significant detector photon pile up, and can cause charge trapping in the CdTe. The alternative is to measure the photons scattered at 900 by a small scattering object, in our case a Lucite cylinder (PMMA). Starting from the scattered spectrum, corrected to dead time and efficiency and calibrated in energy, we made corrections for the scattering angle and coherent scattering (only at low energies). After that, the Klein Nishina cross section for Compton scattering are applied to the spectra, resulting in the primary spectra. For the 120 kVp beam a spectrum of the direct beam was measured and compared to the one obtained it through the scattered spectrum for the same energy. The half value layer of the same spectrum was determined experimentally and through the calculated spectrum, and the agreement was very good. The efficiency curve presents a discontinuity, also observed by other authors and predicted theoretically, that deforms the corrected spectrum. A polynomial fitting was adjusted to the efficiency curve, smoothing the corrected spectra without and not altering its form. The results were satisfactory and they validated the methodology presented by other authors, using different detectors.
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
alineENDJC.pdf (2.36 Mbytes)
Data de Publicação
2008-12-04
 
AVISO: O material descrito abaixo refere-se a trabalhos decorrentes desta tese ou dissertação. O conteúdo desses trabalhos é de inteira responsabilidade do autor da tese ou dissertação.
  • Vieira, A.A., et al. A portable Compton spectrometer for clinical X-ray beams in the energy range 20 150keV [doi:10.1016/j.apradiso.2010.10.011]. Applied Radiation and Isotopes [online], 2011, vol. 69, p. 350-357.
Todos os direitos da tese/dissertação são de seus autores
CeTI-SC/STI
Biblioteca Digital de Teses e Dissertações da USP. Copyright © 2001-2024. Todos os direitos reservados.