Mémoire de Maîtrise
DOI
https://doi.org/10.11606/D.43.2008.tde-26112008-094456
Document
Auteur
Nom complet
Arlene Linke
Adresse Mail
Unité de l'USP
Domain de Connaissance
Date de Soutenance
Editeur
São Paulo, 2008
Directeur
Jury
Yoshimura, Elisabeth Mateus (Président)
Moralles, Mauricio
Tobias, Carmen Cecília Bueno
Titre en portugais
Espectroscopia de raios X utilizando o espalhamento Compton
Mots-clés en portugais
Física
Física experiemental
Física médica
Radiologia
Resumé en portugais
O estudo do espectro de raios X é importante para predizer a qualidade da imagem em sistemas radiográficos. Nesse trabalho usamos um detector semicondutor CdTe para medir os espectros de raios X na faixa de 25 a 120 kVp. Esse detector é pequeno e fácil de ser transportado. As medidas feitas diretamente do tubo de raios X não são aconselháveis devido à alta taxa de fluência que estão acima dos limites de detecção dos detectores usuais e podem causar armadilhamento de carga no detector CdTe. A alternativa é medir os fótons espalhados a 900 a partir de um pequeno objeto espalhador, no nosso caso um cilindro de lucite (PMMA). A partir do espectro espalhado, corrigido para tempo morto e eficiência, e calibrado em energia, efetuamos correções para o ângulo de espalhamento e espalhamento coerente (somente para baixas energias). Só então formalismo de Klein-Nishina para as seções de choque do efeito Compton é aplicado aos espectros, resultando nos espectros primários. Para energia de 120 kVp foi medido um espectro do feixe direto e comparado com o obtido através do espectro espalhado para mesma energia. Para o mesmo feixe, a camada semi-redutora foi avaliada experimentalmente e com o espectro obtido por espalhamento Compton, obtendo-se boa concordância. Observamos na curva de eficiência uma descontinuidade, prevista teoricamente e também observada por outros autores, que deforma o espectro corrigido. Foi aplicado um ajuste polinomial a essa curva, suavizando os espectros obtidos e não alterando sua forma. Os resultados foram satisfatórios e validaram o formalismo apresentado por outros autores, utilizando detectores distintos.
Titre en anglais
X-ray spectroscopy using compton scattering
Mots-clés en anglais
Experimental physics
Medical physics
Physics
Radiology
Resumé en anglais
The analysis of x-ray spectra is important to predict image quality in radiographic systems. In this work we used a semiconductor CdTe detector to measure x-ray spectra with a CdTe detector, in the range from 25 to 120 kVp. The CdTe detector is small and portable. The measurements done directly in the primary X rays beam can be difficult because of the high photon fluence rates that cause significant detector photon pile up, and can cause charge trapping in the CdTe. The alternative is to measure the photons scattered at 900 by a small scattering object, in our case a Lucite cylinder (PMMA). Starting from the scattered spectrum, corrected to dead time and efficiency and calibrated in energy, we made corrections for the scattering angle and coherent scattering (only at low energies). After that, the Klein Nishina cross section for Compton scattering are applied to the spectra, resulting in the primary spectra. For the 120 kVp beam a spectrum of the direct beam was measured and compared to the one obtained it through the scattered spectrum for the same energy. The half value layer of the same spectrum was determined experimentally and through the calculated spectrum, and the agreement was very good. The efficiency curve presents a discontinuity, also observed by other authors and predicted theoretically, that deforms the corrected spectrum. A polynomial fitting was adjusted to the efficiency curve, smoothing the corrected spectra without and not altering its form. The results were satisfactory and they validated the methodology presented by other authors, using different detectors.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
alineENDJC.pdf (2.36 Mbytes)
Date de Publication
2008-12-04
 
AVERTISSEMENT: Le matériau se réfère à des documents provenant de cette thèse ou mémoire. Le contenu de ces documents est la responsabilité de l'auteur de la thèse ou mémoire.
  • Vieira, A.A., et al. A portable Compton spectrometer for clinical X-ray beams in the energy range 20 150keV [doi:10.1016/j.apradiso.2010.10.011]. Applied Radiation and Isotopes [online], 2011, vol. 69, p. 350-357.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.