Master's Dissertation
DOI
https://doi.org/10.11606/D.43.2011.tde-26042012-142823
Document
Author
Full name
Ana Carolina de Magalhães
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2011
Supervisor
Committee
Yoshimura, Elisabeth Mateus (President)
Gesualdi, Marcos Roberto da Rocha
Muramatsu, Mikiya
Title in Portuguese
Estudo da distribuição de luz vermelha e infravermelha em sangue humano diluído para circulação extracorpórea
Keywords in Portuguese
Distribuição de luz em meios espalhadores
Dosimetria
Laser
Abstract in Portuguese
O laser de baixa intensidade (LBI) é utilizado para tratamento em uma série de aplicações, inclusive na diminuição de processos inflamatórios. Assim, poderia ser utilizado para evitar a síndrome da resposta inflamatória sistêmica (SRIS), processo desenvolvido por alguns pacientes que são submetidos à circulação extracorpórea (CEC) procedimento cirúrgico cardíaco. Os objetivos desse trabalho foram estabelecer e validar um arranjo experimental para o estudo da distribuição de luz internamente a líquidos espalhadores e determinar a incerteza associada; utilizar esse arranjo para o estudo da distribuição de luz em sangue, e determinar o melhor comprimento de onda e melhor modo de iluminação para uma possível implementação da aplicação do LBI durante a CEC; e avaliar possíveis mudanças de propriedades reológicas do sangue provocadas pela sua iluminação com laser. Para tal foi utilizado um arranjo com o líquido espalhador Lipovenos PLR dentro de uma cubeta, com utilização de uma fibra ótica para coleta de luz espalhada em diversos pontos dentro do líquido. Lasers de dois comprimentos de onda foram utilizados, 632,8 nm e 820 nm. Os resultados experimentais da distribuição de luz foram comparados com uma simulação pelo método de Monte Carlo já conhecida, o que validou o arranjo proposto, com a incerteza associada aos resultados experimentais de 7%. O arranjo foi utilizado para o estudo da distribuição da luz dentro de sangue diluído para CEC. Além disso, foi avaliada a transmissão da luz pelas paredes dos tubos utilizados nos circuitos da CEC e a distribuição da luz no sangue contido nesses tubos. Um reômetro rotacional com sistema cone-placa foi utilizado para medir características reológicas de duas amostras de sangue diluído, uma iluminada com laser e a outra não iluminada. A luz de 632,8 nm, ao contrário da luz de 820 nm, consegue se distribuir até distâncias maiores do feixe laser, possibilitando que um maior volume de sangue seja tratado. A iluminação do sangue durante a CEC deve ser feita com o tubo de menor diâmetro em quatro pontos ao seu redor, em apenas uma secção do tubo, pois o sangue é mantido circulante. Foram observadas mudanças nas propriedades reológicas do sangue iluminado com o laser de 635nm, as quais devem ser mais bem entendidas para o uso em benefício do paciente submetido a CEC.
Title in English
Study of red and infrared light distribution in diluted blood for cardiopulmonary bupass surgery
Keywords in English
dosimetry
Laser
Light distribution in scattering samples
Abstract in English
Low level laser therapy (LLLT) is a treatment used in several applications, including the reduction of inflammatory processes. It could be used to prevent the systemic inflammatory response syndrome (SIRS), which some patients develop after cardiopulmonary bypass (CPB) surgery applied to solve some heart diseases. The objectives of this study were to set an experimental arrangement to study light distribution inside scattering liquids, and determine the associated uncertainty; to use this arrangement to study light distribution inside blood, in order to implement the LLLT during CPB, and, through this study, to determine the best wavelength and the best way to perform the treatment. Possible changes on rheological properties of blood, caused by illumination with laser, were also evaluated. Lipovenos PLR, a scattering liquid, was contained inside a cuvette and an optical fiber was used to collect the scattered light. Two wavelengths were used: 632.8 nm and 820 nm. The arrangement was validated, with uncertainty of 7%, through comparison between the experimental results and Monte Carlo Method simulation previously performed. This arrangement was used to study light distribution inside blood, diluted to the same conditions of CPB procedure. Light transmission through CPB tubes walls and light distribution in blood inside CPB tubes were also evaluated. A rotational rheometer with a cone-plate system was used to test rheological properties of two blood samples, one illuminated with laser and the other non illuminated Compared to the 820 nm light, the 632.8 nm light is scattered further away from the laser beam, turning possible that a bigger volume of blood could be treated. The blood should be illuminated through the CPB tube with the smallest diameter, in four distinct points around the tube, only in one cross section of this tube, because the blood is kept passing through the tube all the time. Changes in rheological properties of blood were observed on blood illuminated with the 635 nm laser. These changes must be better understood in order to help the patient submitted to CPB.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
dissertacao_final.pdf (7.47 Mbytes)
Publishing Date
2012-05-21
 
WARNING: The material described below relates to works resulting from this thesis or dissertation. The contents of these works are the author's responsibility.
  • A. C. Magalhães, et al. LILT IN CARDIOPULMONARY BYPASS SURGERY ‐ HOW TO DETERMINATE THE OPTICAL PROPERTIES AND LIGHT DISTRIBUTION IN BLOOD?. In The World Association of Laser Therapy Conference 2008, Sun City, 2008. WALT 2008 - Book of Abstracts., 2008. Abstract.
  • Sousa, M V P, et al. Light distribution in phantoms. In International Conference on Medical Physics, Porto Alegre, 2011. Proceedings of the International Conference on Medical Physics., 2011. Abstract.
  • A. L. O. Ramos, et al. IMPROVING LOW INTENSITY LASER THERAPY WITH MONTE CARLO SIMULATION. In The World Association of Laser Therapy Conference 2008, Sun City, 2008. WALT 2008 - Book of Abstracts., 2008. Abstract.
  • Sousa, M V P, et al. Phantoms of fingers with various tones of skin for LLLT dosimetry. In SPIE PHOTONICS WEST BIOS, 2011, SAN FRANCISCO, 2011. SPIE PROCEEDINGS. : Society of Photo-Optical Instrumentation Engineers (SPIE)., 2011.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.