Master's Dissertation
DOI
https://doi.org/10.11606/D.43.2007.tde-09052007-195209
Document
Author
Full name
Marcelo Scolaro Morlotti
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2007
Supervisor
Committee
Yoshimura, Elisabeth Mateus (President)
Caldas, Linda Viola Ehlin
Nicolucci, Patrícia
Title in Portuguese
Verificação de tratamentos radioterápicos diversos com dosimetria termoluminescente
Keywords in Portuguese
Dosimetria Termoluminescente
Física Médica
IMRT
Radioterapia 3D
TLD100
Abstract in Portuguese
As técnicas radioterápicas evoluíram muito nas últimas décadas, destacando-se a utilização de feixes segmentados por colimadores de multifolhas, os quais permitem entregar maior quantidade de dose com maior precisão espacial. Os sistemas de planejamento em radioterapia (TPS), entretanto, são em grande parte fechados, com acesso restrito aos parâmetros de cálculo de dose. Desta forma, o presente trabalho teve como principal objetivo comparar valores de dose envolvidas em tratamentos radioterápicos, medidos com a técnica de dosimetria termoluminescente (TL), usando um objeto simulador antropomórfico (Alderson Rando Phantom - ARP) como paciente, com valores de referência extraídos dos TPS em simulações realistas de tratamento. O objeto simulador continha dosímetros TL de LiF (TLD100) posicionados na pelve, no tórax, na cabeça e no pescoço, a fim de que fossem monitoradas regiões com dose homogênea (como no volume alvo planejado - PTV), regiões com gradiente de dose e regiões afastadas do PTV. As irradiações foram feitas usando técnicas de radioterapia conformacional 3D e radioterapia por modulação da intensidade do feixe (IMRT). Quatro TPS foram utilizados: CadPlan, ECLIPSE, Helax-TMS e KonRad. Em algumas situações utilizaram-se algoritmos de cálculo implementados pelos fabricantes nos TPS; dois deles baseados em correção de heterogeneidades (Batho e pencil beam) e um deles baseado em convolução e superposição de feixes simulados (collapsed cone). Nos casos de irradiações com IMRT os algoritmos foram sempre usados. Paralelamente ao uso do ARP, utilizou-se um objeto simulador com água com objetivo de verificar o comportamento dos dosímetros TL expostos a feixes de fótons de alta energia, em comparação com câmaras de ionização cilíndricas do tipo Farmer (volume interno de 0,6 cm3). Além disso, esse objeto simulador possibilitou garantir a qualidade de suportes desenvolvidos para a câmara e para os dosímetros, nos casos de medidas em água. A irradiação da pelve do objeto simulador ARP, região com poucas heterogeneidades, revelou que o comportamento do material que o constitui é equivalente ao da água e, por isso, pode ser usado na simulação de tecido humano em feixes de radioterapia. Simulações de tratamentos revelaram compatibilidade entre valores medidos e planejados no interior do PTV, com discrepâncias menores que 2%; valores que estão dentro dos limites estabelecidos pelo ICRU62 (+7%/-5%, 2?). Nas regiões afastadas do PTV, todos os códigos de planejamento apresentaram discrepâncias relativamente grandes, evidenciando limitações no cálculo de componentes secondárias. Apesar dos algoritmos de cálculo aumentarem a concordância entre doses medidas e planejadas, não foi possível identificar diferenças entre os algoritmos pencil beam e collapsed cone. Em regiões onde havia alto gradiente de dose, as discrepâncias se tornam maiores devido à dificuldade em posicionar os dosímetros no mesmo ponto em que o cálculo é feito. Sistematicamente, observou-se que a técnica para radioterapia 3D apresenta valores de dose dentro dos limites preestabelecidos, enquanto IMRT mostra valores com maior exatidão.
Title in English
Assessment of radiotherapic treatments with thermoluminescent dosimetry
Keywords in English
3D radiotherapy
IMRT
Medical Physics
Thermoluminscent dosimetry
TLD100
Abstract in English
The last decades noticed a massive improvement in radiotherapy techniques and the use of segmented beams produced by multileaf collimators. However, the dose radiation therapy planning systems (TPS) in use are characterized by the restriction of access to the calculation algorithms. The aim of this work was to compare dose values measured in an anthropomorphic phantom (Alderson Rando Phantom - ARP) to the reference dose values obtained from the TPS in real case simulations of radiotherapy treatments. Thermoluminescent dosimetry (TLD) technique was used to evaluate the doses. LiF dosimeters (TLD100) were positioned on the phantom pelvis, thorax, head and neck at homogeneous dose regions, as the Planned Target Volume (PTV), gradient dose regions, and areas far from PTV. The doses were delivered using two techniques, 3D conformal radiotherapy and Intensity Modulated Radiation Therapy (IMRT). Four TPS were used: CadPlan, ECLIPSE, Helax-TMS and KonRad. In several situations, calculation algorithms implemented in these planning systems were employed to take heterogeneities into account: two of them were correction-based algorithms (Batho and Pencil Beam) and one of them based in beam convolution-superposition (Collapsed Cone). Furthermore, a liquid water phantom was used to compare the TLD behavior to the Farmer thimble chamber (0,6 cc internal volume) results when exposed to high-energy photon beams. It was also possible to verify the quality of the PMMA supports that wer used in the calibration of the dosimeters on clinical beam, obtaining reliable results. Both the ARP pelvis and a water phantom showed similar behavior under irradiation, indicating that, in highly homogenous regions, the ARP material can be used to simulate human soft tissues under radiotherapy treatment. Inside the PTV, the dosimetry performed in the ARP, showed compatibility between measured and planned dose values, with discrepancies smaller than 2%, which are within the ICRU62 fixed limits (+7%/-5%, 2?). When the algorithms were used, a better agreement between the experimental and planned doses was achieved, but it was not possible to discriminate the Pencil Beam and Collapsed Cone algorithms. In regions with large dose gradients, the discrepancies between experimental and planned dose values are higher as the difficulties the position of the dosimeters are more critical. The measured doses, when the 3D radiotherapy technique was used, were within the ICRU62 pre-established limits whereas the IMRT technique provider more accurate values.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2007-05-29
 
WARNING: The material described below relates to works resulting from this thesis or dissertation. The contents of these works are the author's responsibility.
  • MORLOTTI, M. S., and YOSHIMURA, E. M. Internal dose measurements in photon beam radiotherapy using an anthropomorphic phantom (ARP) with TLD-100 dosimeters. In 15th International Conference on Solid State Dosimetry, Delft, Holanda, 2007. Book of Abstracts., 2007. Abstract.
  • MORLOTTI, M. S., FURNARI, L., e YOSHIMURA, E. M. Verificação de tratamentos radioterápicos diversoscom dosimetria termoluminescente (TL). In X Congresso Brasileiro De Física Médica, Salvador - Bahia, 2005. X Congresso Brasileiro De Física Médica - Resumos., 2005. Resumo.