Disertación de Maestría
DOI
https://doi.org/10.11606/D.43.2001.tde-02082013-163614
Documento
Autor
Nombre completo
Marcia de Carvalho Silva
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2001
Director
Tribunal
Herdade, Silvio Bruni (Presidente)
Caldas, Linda Viola Ehlin
Yoshimura, Elisabeth Mateus
Título en portugués
Espectroscopia de raios X na faixa de energia de 5 a 200 keV, utilizando fotodiodos PIN de silício
Palabras clave en portugués
Espectroscopia
Fotodiodos
Radiodiagnóstico
Raios-X
Resumen en portugués
O conhecimento da distribuição espectral da radiação X emitida por unidades de radiodiagnóstico fornece importantes informações, que podem ser aplicadas aos programas de Garantia de Qualidade e de Proteção Radiológica. Detectores que utilizam fotodiodos PIN de Si como ponta de prova vêm sendo utilizados nos últimos anos devido, principalmente, ao seu baixo custo e por trabalharem a temperatura ambiente, não precisando dos tanques de nitrogênio líquido necessários em detectores de Ge e Si(Li). Embora, a princípio, os fotodiodos PIN tenham uma pior resolução, quando comparada com os detectores de Ge, ela é suficientemente boa para a maior parte das aplicações em radiodiagnóstico. Além disso, o desenvolvimento de sistemas refrigerados termoelétricamente e de pré-amplificadores de baixo ruído têm feito com que os fotodiodos PIN cheguem a resoluções comparáveis às do Ge. Este trabalho visa estabelecer as propriedades de detectores constituídos por fotodiodos PIN de Si e desenvolver o processo de correção dos espectros brutos para a eficiência do detector e outros fatores pertinentes, a fim de se obter o espectro real emitido por equipamentos de radiodiagnóstico. Espectros de radiação de equipamentos de radiodiagnóstico obtidos experimentalmente foram comparados com espectros teóricos calculados a partir de um modelo semi-empírico. Além disso, foram obtidos espectros de radiação emitidos por unidades de mamografia, espalhados por um objeto simulador de mama. Como aplicação direta dos espectros medidos experimentalmente, foram desenvolvidas metodologias para a determinação da tensão aceleradora de tubos de Raios X (kVp), para fins de calibração de medidores de kVp, e da camada semi-redutora (CSR) dos feixes de interesse.
Título en inglés
X-ray spectroscopy in the energy range from 5 to 200 keV, using silicon PIN photodiodes
Palabras clave en inglés
Photodiodes
Radiology
Spectroscopy
X-rays
Resumen en inglés
Knowledge of the spectral distribution of radiation emitted by radiodiagnostic units provides important information, which can be applied to the Quality Assurance and Radiological Protection programs. Detectors which utilize Silicon PIN Photodiodes have been increasingly used in recent years, due to their low cost and being worked with at room temperature, the liquid nitrogen containers used with Ge and Si(Li) detectors not being necessary. Although, initially, the PIN photodiodes have a poorer energy resolution, as compared with Ge detectors, it's good enough for most purposes. Moreover, the development of thermoelectrically cooled systems and low noise level pre-amplifiers have resulted in PIN photodiodes achieving comparable resolutions to Ge ones. This work seeks to establish the properties of detectors made with Silicon PIN photodiodes and to develop the correctional process of the raw specffa both for the efficiency of the detector and other relevant factors, with the aim of obtaining the true spectrum emitted by radiodiagnostic equipment. The radiation spectra of radiodiagnostic equipment experimentally obtained were compared with theoretical spectra calculated from a semi-empirical model. Furthermore, mammography x-ray spectra scattered by a breast simulator were obtained. As a direct application of these experimentally measured spectra, methodologies were developed to determine the peak kilovoltage (kvp) of X-Ray tubes, to be able to calibrate kVp meters, and the half-value layer (HW) of the beams of interest.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
PS32639SILVA.pdf (4.35 Mbytes)
Fecha de Publicación
2013-08-02
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.