Disertación de Maestría
DOI
https://doi.org/10.11606/D.43.1990.tde-07052014-181341
Documento
Autor
Nombre completo
Eduardo Luiz Augusto Macchione
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 1990
Director
Tribunal
Dietzsch, Olacio (Presidente)
Cusatis, Cesar
Ueta, Nobuko
Título en portugués
Detector de raios-X sensível à posição.
Palabras clave en portugués
Detector proporcional sensível à posição
Instrumentação nuclear
Raios-x
Resumen en portugués
Um contador a gás, com multifios, sensível à posição, para a detecção de raios X foi construído utilizando linhas de atraso integradas para leitura da posição. Seis linhas de atraso (50ns de atraso cada, frequência de corte de 40MHz) cobrem um comprimento sensível total de 150mm, possuindo uma razão do atraso total pelo tempo de subida dos sinais na linha adequadas para a detecção em posição raios X de 5,9keV do ANTIPOT. 55 Fe com alta resolução. Testes usando a forneceram uma não-linearidade linha de integral máxima menor que 0,1% e não-linearidade diferencial máxima de ±4,0%, com o detector utilizando a mistura 700torr. Testes similares foram feitos usando de Ar-CH IND 4 (90%-10%) a a linha de 8,04keV de um tubo de raios X de Cu. Uma resolução total de 330m, e as mesmas não-linearidades integral e diferencial foram obtidas. Com a finalidade de t.est.ar o comportamento do detector sob condições experimentais reais, duas diferentes amostras foram analisadas na região de espalhamento em baixo ângulo. As amostras analisadas foram o carvão vítreo e o dodecil sulfato de sódio (SLS). O feixe de raios X foi colimado por um conjunto de fendas, com 50,um cada uma delas e separadas de 98mm. A distância entre a amostra e o detector era de 404,5mm, fazendo com que a região sensível do detector cobrisse um ângulo de espalhamento de ±12 °. Foram obtidos um raio de giro de 22 Å para o carvão vítreo e um dos parâmetros da rede cristalina c de 39,1 Å para o SLS. Estes resultados estão de acordo com trabalhos anteriores obtidos com técnicas tradicionais de maior dispêndio de tempo que utilizam detectores de Nal(T1) e chapas fotográficas em uma câmara de Laue.
Título en inglés
X-rays detector sensitive to the position.
Palabras clave en inglés
Nuclear Instrumentation
proportional detector sensitive to the position
X-rays
Resumen en inglés
A multiware position sensitive gas counter for X-Ray detection was developed in our laboratory making use or commercial delay-lines for position sensing. Six delay-line chips (50ns delay each 40Mhz cut-off frequency) cover a total sensitive length of 190mm leading to a delay-risetime ratio that allows for a high-resolution position detection. Tests using the 5,9keV X-Ray line from a 55Fe source an integral linearity better than 0,1% and a maximum differential linearity or ±4,0% were obtained operating the detector with an Ar-CH 4 (90%-10%) gas mixture at 700torr. Similar tests were performed using the 8,04keV line from a Cu X-Ray tube. A total resolution or 330m, and the same integral and differential linearities were obtained. In order to check the detector performance under usual experimental conditions two different samples were analyzed in the low scattering-angle region. The samples analyzed were vitreous carbon and sodium dodecyl sulfate (SLS). The X-Rays were collimated by a set or slits each 50m wide separated by 98mm. The distance between the sample and the detector was 404,5mm, so that the detector covered a scattering angle of ±12°. A radius of gyration of 22Å for vitreous carbon and a crystallographic parameter of 39,1Å for SLS were obtained. These values are in good agreement with those obtained with the more time-consuming traditional techniques making use of a rotating NaI(T1) detector and photographic plates with a Laue camera.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Macchione.pdf (7.24 Mbytes)
Fecha de Publicación
2014-05-16
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.