Doctoral Thesis
DOI
https://doi.org/10.11606/T.43.1998.tde-04122013-145919
Document
Author
Full name
Eduardo Luiz Augusto Macchione
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 1998
Supervisor
Committee
Dietzsch, Olacio (President)
Cruz, Manoel Tiago Freitas da
Souza, Gerardo Gerson Bezerra de
Ueta, Nobuko
Vieira Junior, Nilson Dias
Title in Portuguese
Espectrômetro de Massa por Tempo de Vôo para Análise de Macromoléculas
Keywords in Portuguese
Física nuclear
Instrumentação
Abstract in Portuguese
Um espectrômetro de massa por tempo de vôo foi projetado e construído. A técnica de Plasma Desorption é empregada para provocar a dessorção e ionização das moléculas da amostra por colisões com os produtos de fissão emitidos de uma fonte radioativa de ANTPOT.252 Cf. O espectrômetro encontra-se instalado no interior de uma câmara de vácuo, que permite a sua operação a pressões da ordem de 10 POT.-7 torr. Uma pré-câmara é utilizada para a manipulação e troca de até três amostras simultaneamente, sem a necessidade da abertura da câmara principal para a atmosfera, reduzindo o tempo de recuperação da pressão do sistema para início das análises. No tubo de vôo, que possui uma extensão de 111 cm, encontra-se instalado um guia eletrostático de partículas, usado para aumentar a transmissão dos íons secundários até o sistema de detecção. Medidas foram feitas utilizando-se amostras-padrão para a caracterização do instrumento. Estudos também foram realizados sobre o comportamento da resolução e da eficiência de transmissão do aparelho em função entre 1600 e 1800 para íons com massas em torno de 900 u.m.a.
Title in English
Time-of-flight mass spectrometer for analysis of macromolecules
Keywords in English
Instrumentation
Nuclear physics
Abstract in English
A time-of-flight mass spectrometer was projected and constructed. The Plasma Desorption technique is employed to desorb and ionize the molecules sample by collisions with fission fragments emitted from ANTPOT.252 Cf source. The spectrometer is placed inside a vacuum chamber, allowing operation at a pressure around of 10 POT.-7 torr. A pre-chamber is used to insert three samples simultaneously into the mass spectrometer, without the need of opening the main chamber to the atmosphere, reducing the recover time of the pressure system for the beginning of analyzes. Inside the flight tube (111 cm length) was installed an electrostatic particle guide, used to increase the secondary ion transmission to the detection system. Measurements using standard samples were done to characterize the apparatus performance. The instrument resolution and transmission efficiency were analyzed with respect to the secondary ion acceleration voltage. The mass resolution of m/m from 1600 to 1800 for rnass ions around 900 u.m.a. was obtained.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2014-02-20
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.