Titre en portugais
Estudo experimental de escoamento pulsátil através de biopróteses valvulares cardíacas mitrais a partir do desenvolvimento de um simulador do lado esquerdo do coração.
Mots-clés en portugais
Bioprótese
Hemodinâmica
Simulador cardíaco
Valva mitral
Velocimetria
Resumé en portugais
Uma vez que a maioria das complicações relacionadas ao funcionamento das próteses de válvulas cárdicas é devida aos distúrbios de escoamento, a sua caracterização hidrodinâmica é um auxílio útil no projeto de novas válvulas. Simulações do escoamento pulsátil através de próteses de válvulas cardíacas começaram há cerca de 60 anos, por meio do desenvolvimento de diferentes simuladores cardíacos, melhorando a interpretação dos resultados clínicos. Para o presente estudo, um simulador cardíaco foi desenvolvido completamente na Escola Politécnica da Universidade de São Paulo, no Departamento de Engenharia Mecânica, para estudar o fluxo através de próteses mitrais e aórticas. A sua concepção é baseada na geometria e na fisiologia do lado esquerdo do coração humano, no estado da arte desses simuladores e na norma ISO 584:2010. Neste trabalho, o simulador cardíaco desenvolvido foi aplicado para estudar experimentalmente o fluxo pulsátil através de próteses cardíacas mitrais biológicas de pericárdio bovino. Para tanto, após a validação do simulador para algumas freqüências cardíacas, as biopróteses foram submetidas a ensaios de velocimetria não invasiva, por meio de um equipamento Particle Image Velocimetry (PIV). Os experimentos, que se concentraram principalmente logo após o começo da diástole, mostraram qualitativamente uma diferença significativa no escoamento intraventricular em função de dois posicionamentos das biopróteses, defasadas entre si de 180º no sentido do seu eixo longitudinal. Além disso, produziram significativas mudanças no escoamento próximo à prótese aórtica considerada. Com a continuidade deste estudo, os resultados poderão induzir a um dado procedimento clínico.
Titre en anglais
Experimental study pulsatile flow through mitral heart bioprosthesis based on the development of a left human heart cardiac simulator.
Mots-clés en anglais
Bioprosthesis
Cardiac simulator
Hemodynamics
Mitral valve
Velocimetry
Resumé en anglais
Since most complications related to the operation of prosthetic heart valves are due to disturbances of flow, their hydrodynamic characterization is a useful aid in the design of new prostheses. Simulations of pulsatile flow in cardiac prostheses began nearly 60 years ago, through the development of different mock human circulatory systems, improving the clinical results interpretation. A complete design of a cardiac simulator was developed at Polytechnic School of USP, Mechanical Engineering Department, to study mitral and aortic prosthetic heart valves. The simulator design is based on the both geometry and physiology of the left side of the human heart, on state-of-art studies and also considering the ISO 584:2010 standard. The cardiac simulator was applied to pulsatile flow testing of bovine pericardium mitral prosthetic valves. For this, after the cardiac simulator validation for some heart rates, the bioprostheses were assayed by noninvasive velocimetry technics, i.e., a Particle Image Velocimetry (PIV) system. The testing, which mainly were focused on just after the beginning of diastole, qualitatively have showed a significant difference in the intraventricular flow according two different positions for the bioprostheses, dephased of 180° in the direction of their longitudinal axis. Moreover, they produced significant changes in flow close to the aortic prosthesis considered. Future works may lead to a particular medical procedure.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
TESE_OVANDIR_BAZAN.pdf (10.95 Mbytes)
Date de Publication
2015-01-07
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.