Title in Portuguese
Análise do comportamento térmico da estrutura intra-ocular humana durante procedimento cirúrgico de remoção de catarata.
Keywords in Portuguese
Análise térmica
Catarata
Estrutura intra-ocular
Resfriamento do fluido de irrigação
Abstract in Portuguese
O presente trabalho apresenta uma análise do comportamento térmico da estrutura intra-ocular humana durante a cirurgia de remoção de catarata. O procedimento cirúrgico em estudo, conhecido por facoemulsificação, baseia-se em destruir o cristalino opaco, removê-lo da estrutura intra-ocular e substituí-lo por uma lente artificial. A destruição do cristalino é realizada por aplicação de energia ultra-sônica, resultante da vibração do cristal piezoelétrico existente na ponteira do equipamento cirúrgico. O campo acústico imposto à estrutura intra-ocular aquece a mesma, elevando sua temperatura até valores críticos. Pelo fato da córnea poder atingir valores entre 40oC e 115oC durante este procedimento, inúmeras complicações pós-operatórias podem ser diagnosticadas, tais como: danos ao estroma e endotélio da córnea, formação de fístula, astigmatismo residual e impossibilidade de autoselamento da incisão, sendo necessário suturar a esclera. Para o estudo deste superaquecimento identificado como crítico, foi realizada a modelagem do comportamento térmico da estrutura intra-ocular humana durante a facoemulsificação. Foram desenvolvidos modelos para quantificar o campo acústico imposto à estrutura pelo equipamento, bem como para avaliar o aquecimento e o resfriamento da estrutura ocular. Dos resultados obtidos a partir destes modelos, o equilíbrio térmico nos tecidos da córnea forneceu o resfriamento necessário a ser imposto pelo fluido de irrigação (6,62 W), a fim de se manterem os níveis normais de temperatura na estrutura (30ºC). Enquanto na literatura estudos têm sido apresentados no sentido de melhorar a facoemulsificação ou mesmo desenvolver novas técnicas, mais complexas, em busca de menor prejuízo à integridade da estrutura, a análise conjunta entre medicina e engenharia desenvolvida neste trabalho apresentou como resultado a proposta de resfriamento do fluido utilizado na irrigação, sendo esta uma solução simples e prática. Para permitir o resfriamento do fluido, dimensionou-se um sistema de troca de calor acessório ao equipamento utilizando o Efeito Peltier, por permitir a fabricação de resfriadores miniaturizados, compactos e eficientes.
Title in English
Thermal behavior analysis of human intra-ocular structure during surgical procedure of cataract removing.
Keywords in English
Cataract
Intra-ocular structure
Irrigation fluid cooling
Thermal analysis
Abstract in English
This research presents an analysis of the thermal behavior of the human intra-ocular structure during surgery of cataracts removing. The surgical procedure under study, known as phacoemulsification, is based on destructing the lens with cataracts, removing it from the intra-ocular structure and changing it into artificial lens. The destruction of the lens is achieved by using ultrasound energy, resulting from the vibration of a piezoelectric crystal existent in the surgical equipment. The acoustic beam incident in the intra-ocular structure heats it, increasing its temperature up to critical values. As the corneal structure can achieve from 40oC to 115oC during this procedure, many postoperative complications can happen, such as: damage to the corneal stroma and endothelium, fistula formation, induced astigmatism and inability to close the incision, necessitating a scleral patch graft. For the studies regarding this over heating identified as critical, a model of the thermal behavior of the human intraocular structure during phacoemulsification was created. Models were developed to quantify the acoustic beam incident in the structure by the equipment, as well as models to valuate the heating and cooling of the intra-ocular structure. With the results from these models, the thermal equilibrium on the corneal tissues gave the necessary cooling to be incident by the irrigation fluid (6,62 W) so that the normal temperature levels could be kept on the structure (30ºC). While in the literature studies have been presented on the way to improve phacoemulsification or even develop new techniques, more complex, aiming to have less prejudice to the integrity of the structure, the analysis together involving medicine and engineer developed within this research presented as result a proposal of cooling the irrigation fluid, what seemed to be a simple and practical solution. To allow this cooling, it was dimensioned a heat changing system as an accessory to the surgical equipment using the Peltier effect, as it allows the construction of miniaturized, compact and efficient coolers.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.