Disertación de Maestría
DOI
https://doi.org/10.11606/D.3.2014.tde-30122014-110751
Documento
Autor
Nombre completo
Stelvio Henrique Ignacio Barboza
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2014
Director
Tribunal
Kofuji, Sergio Takeo (Presidente)
Minami, Mário
Van Noije, Wilhelmus Adrianus Maria
Título en portugués
Tomógrafo em nível de simulação utilizando micro-ondas em banda ultra larga (UWB) com transmissor em tecnologia CMOS para detecção precoce de câncer de mama.
Palabras clave en portugués
Detector de imagens
Micro-ondas
Microeletrônica
Tomógrafo
Resumen en portugués
O sistema desenvolvido obteve boa resposta na detecção de modelos numéricos de tumores com dimensões a partir de 5mm, representada na localização adequada e determinação do tamanho obtidos por meio de simulações envolvendo os modelos dos blocos especificados. Como objetivo principal do trabalho será apresentado o projeto, fabricação e resultados de testes de um circuito integrado gerador de pulsos com o formato da derivada de quinta ordem do pulso de gaussiano (transmissor UWB) fabricado utilizando a tecnologia IBM 0.18 CMOS. Os blocos principais que formam o circuito gerador de pulso são: circuito gerador de onda quadrada, gerador de atrasos, detector de fase e etapa de saída (formador de pulso). O gerador de onda quadrada foi implementado a partir de um buffer de RF com um inversor na saída com casamento de impedância. O gerador de atrasos foi implementado a partir de uma cascata de inversores. O circuito detector de fase é composto por bloco n- dinâmico , n-latch e inversor estático para forma pulsos em alta velocidade. As dimensões dos transistores foram definidas de modo a obter característica adequada de um pulso Gaussiano de 5ª ordem, considerando especificações exigidas de Sistema de Detecção de Câncer de Mama. Foi implementado o leiaute em full Custom com dimensões mínimas da tecnologia. Cinco chips diferentes foram testados. E os valores da fonte de alimentação foram variadas em 1,62V, 1,80V e 1.98, então foram medidos os valores de saída pico a pico e largura de pulso para cada chip. O consumo de energia medido foi de 244 uW, e a amplitude do pulso de saída 115,2 mV pico a pico e largura de pulso de 407,8 ps com um sinal de entrada senoidal de amplitude 806mVp à 100 MHz . O pulso gerado pelo gerador de pulso resultou em uma PSD (Power Spectral Density) com largura de banda de 0,6 GHz a 7,8 GHz, que é adequado para aplicações de UWB para detecção do câncer de mama.
Título en inglés
Tomograph in simulation level using microwave in ultra wide band with transmitter in CMOS technology for early breast cancer detection.
Palabras clave en inglés
CT scanner
Detector images
Microelectronics
Microwave
Resumen en inglés
As a result for detecting the numeric representation, the system could identify tumors from 5mm of extent with adequate localization, as well size determination. The primary goal of this work, therefore, is to bring out the project, manufacturing process and achieved results of tests regarding an integrated circuit for generating pulses which are shaped as the derivative of fifth order of the Gaussian pulse (UWMB transmitter) using the UWB 0.18 CMOS. The pulse generator circuit is composed by: circuit for generating square waves, delay generator, phase detector and output stage. The generator of square wave was implemented from one buffer of rf, with an inverter in the output and impedance matching. The delay generator was implemented from one cascade of inverters. The circuit for detecting the stages is assembled with n block dynamic, n-latch and static inverter for quickly generating pulses (high speed pulse generation). The dimensions of the transistors were defined in order to obtain the adequate characteristics of one Gaussian pulse of 5th order, considering the required specification of the Detection System for Cancer of Breast. It was implemented using the full Custom layout, taking into account the minimum dimensions for such technology. Five different chips were tested. The values of the source energy varied among 1,62V, 1,80V and 1,98V, being later measured the output values, peak to peak, as well the pulse width for each chip. The measured energy consumption was 244 uW, the amplitude of the output pulse was 115.2 mV peak to peak, and the pulse width was 407,8ps with sinusoidal input signal of 806mVp amplitude at 100MHz. As a result, it was obtained a PSD (Power Spectral Density) with band width of 0,6GHz to 7,8GHz from the pulse generator, which is quite adequate for UWB applications for detecting the breast cancer.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2015-01-07
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.