• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Doctoral Thesis
DOI
https://doi.org/10.11606/T.3.2018.tde-18072018-135559
Document
Author
Full name
Raphael Garcia Moreira
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2018
Supervisor
Committee
Santos Filho, Sebastião Gomes dos (President)
Mansano, Ronaldo Domingues
Santos, Edson Tafeli Carneiro dos
Silva, Maria Lucia Pereira da
Siqueira, Cesareo de La Rosa
Title in Portuguese
Fabricação e modelagem de uma nova geometria para espectrometria de mobilidade iônica de tolueno, propanol e água com elevada relação sinal-ruído.
Keywords in Portuguese
Descarga elétrica
Espectrometria
Espectrômetro
Ionização de gases
Relação entre sinal e ruído
Abstract in Portuguese
Neste trabalho é apresentada uma nova geometria e os critérios de projeto para dispositivos de espectrometria de mobilidade iônica, com objetivo de aumentar a relação entre sinal e ruído detectados pelos eletrômetros que compõem a região de detecção, quando associado à ionização baseada em descargas elétricas que tipicamente emitem maior interferência eletromagnética em relação a outros tipos de ionização. A geometria e os critérios de projeto foram estabelecidos com o auxílio do simulador de trajetórias iônicas SIMION versão 8.1 e comparada com espectrômetros de mobilidade iônica do estado-da-arte. Os resultados das simulações dos dispositivos foram avaliados por meio dos testes estatísticos não-paramétricos de Iman-Davenport e Holm. O resultado indicou que a geometria proposta apresentou compatibilidade com as melhores características dos espectrômetros atuais com confiança estatística a 95 %. Na sequência a geometria proposta foi simulada, fabricada e testada para detecção dos íons de: tolueno - C7H8 (C5H5+, C6H5+, C7H7+), propanol - C3H8O (CH3+, CH3O+, C2H3+, C2H4O+, C2H5O+, C3H6+, C3H7+, C3H7O+) e água - H2O (H+, HO+, H2+, O+). Os picos de corrente iônica detectada nos eletrômetros indicaram coerência com as trajetórias simuladas para estes conjuntos de íons. Além disso, os resultados experimentais apresentaram um aumento na relação entre sinal e ruído para a nova geometria proposta que foi validada pelo teste de comparação pareada não-paramétrico de Wilcoxon a 95 % de confiança estatística.
Title in English
Fabrication and modeling of a new geometry for ion mobility spectrometry of toluene propanol and water with high signal-to-noise ratio.
Keywords in English
Discharge ionization
Ion mobility spectrometry
Signal-to-noise ratio
Spectrometer
Abstract in English
In this work, a new geometry and a design criteria for ion mobility spectrometry devices was presented, aiming to increase the signal-to-noise ratio detected by electrodes of the detection region, when associated with ionization based on electrical discharges. The new geometry and the design criteria were established in the SIMION version 8.1 (an ionic trajectory simulator) and compared with the state-of-art ion mobility spectrometers. The results from device simulations were evaluated using the non-parametric statistical test of Iman-Davenport and Holm. The result indicated that the proposed design criteria presented compatibility with the best characteristics of current spectrometers with statistical evidence at 95%. The proposed geometry was simulated, fabricated and tested for the following solvents: toluene - C7H8 (C5H5+, C6H5+, C7H7+), propanol - C3H8O (CH3+, CH3O+, C2H3+, C2H4O+, C2H5O+, C3H6+, C3H7+, C3H7O+) and water - H2O (H+, HO+, H2+, O+). The peaks of ion current detected in the electrometers occurred at approximate positions of the simulated trajectories for these sets of ions. In addition, the experimental data showed an increase in the signal-to-noise ratio for the proposed new geometry that was validated by the test statistic of Wilcoxon signed rank at 95% statistical confidence.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2018-07-26
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.