• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Disertación de Maestría
DOI
https://doi.org/10.11606/D.25.2011.tde-16082011-085632
Documento
Autor
Nombre completo
Samira Salmeron
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
Bauru, 2011
Director
Tribunal
Rezende, Maria Lucia Rubo de (Presidente)
Damante, Carla Andreotti
Souza, Sergio Luis Scombatti de
Título en portugués
Efeitos do laser em baixa intensidade, da terapia fotodinâmica e do azul de toluidina O na descontaminação de superfícies de implantes metálicos: estudo microscópico em subcutâneo de ratos
Palabras clave en portugués
Azul de toluidina O
Descontaminação
Implante dentário
Terapia a laser
Terapia fotodinâmica
Resumen en portugués
A peri-implantite acomete um número crescente de indivíduos e os protocolos de tratamento objetivam descontaminar as superfícies dos implantes e torná-las novamente osseointegráveis. Porém, ainda não foi estabelecido um padrão ouro de tratamento para peri-implantite. Este estudo testa o poder de descontaminação do laser em baixa intensidade (LBI), da terapia fotodinâmica (PDT) e do azul de toluidina O (TBO). Sobre discos de titânio de 1,5 mm de espessura e 4,0 mm de diâmetro de superfícies lisas ou rugosas foi permitida a deposição natural de biofilme microbiano por 7 dias após os quais os discos foram divididos em grupos experimentais (15 discos por grupo), de acordo com o método de descontaminação: grupo LBI (laser de InGaAlP de 660 nm, 30 mW, 45 J/cm2 no modo continuo por 30 s); grupo PDT; grupo TBO (aplicação de TBO por 60 s); grupo controle positivo (C) sem tratamento e grupo controle negativo (NC) estéril. Os discos foram implantados em tecido conjuntivo subcutâneo de ratos e, após 7, 28 e 84 dias, foram obtidas 5 biópsias de cada grupo para análise em microscopia óptica. Implantes lisos e rugosos não diferiram com relação ao grau de fibrosamento e severidade do infiltrado inflamatório, mas a área do tecido reacional perimaterial foi maior nos rugosos (2,6 ± 3,7 x 106 µm2) do que nos lisos (1,9 ± 2,6 x 106 µm2). O grupo C foi o que apresentou menor grau de fibrosamento do tecido reacional (1), mas somente houve diferença estatisticamente significante (p = 0,0230) entre os grupos C e NC. A severidade do infiltrado inflamatório somente diferiu significantemente entre os grupos aos 7 dias, quando o grupo NC apresentou o menor escore (2) em comparação aos demais grupos (3). Com relação à área do tecido reacional perimaterial, só houve diferenças entre os grupos aos 7 dias, quando o grupo C apresentou maior área que os demais grupos (9,11 ± 2,10 x 106 µm2), porém, sem diferença estatisticamente significante em comparação aos grupos LBI e TBO. A área apresentada pelo grupo PDT (4,34 ± 1,49 x 106 µm2) só foi inferior ao grupo NC (0,65 ± 0,23 x 106 µm2). A espessura do tecido reacional não confirmou a diferença entre os discos de superfície lisa e os de superfície rugosa (p = 0,2801), mas confirmou todas as demais comparações relativas à área, sendo que o grupo NC apresentou menor espessura que os demais somente aos 7 dias. O grupo NC apresentou menor espessura de tecido reacional do que os demais grupos, mas essa diferença foi estatisticamente significante somente aos 7 dias. Os demais grupos não diferiram entre si quanto a esse parâmetro. Concluiu-se que implantes rugosos ou lisos contaminados por biofilme microbiano e tratados com LBI, PDT e TBO não diferem entre si quanto aos parâmetros inflamatórios teciduais, parecendo haver leve superioridade da PDT sobre os demais tratamentos apenas no período mais precoce de observação. Porém, com o passar do tempo, todos os tratamentos se equiparam a implantes estéreis quanto à reação tecidual perimaterial.
Título en inglés
Effect of low intensity laser, photodynamic therapy and toluidine blue O the on decontamination of metallic implants surfaces: microscopic study in rats' subcutaneous tissue
Palabras clave en inglés
Decontamination
Dental implantation
Laser therapy
Photodynamic therapy
Toluidine blue O
Resumen en inglés
The peri-implantitis is a disease that is increasing in a number of individuals and the protocols of treatment has intended to decontaminate the implant surfaces and makes this biocompatible again. However, until this moment, a gold standard for peri-implantitis treatment was not established. The present study tests the power of decontamination of low intensity laser (LBI), photodynamic therapy (PDT) and toluidine blue O (TBO). The natural deposition of microbian biofilm was allowed on titanium discs with 1,5 mm of thickness and 4,0 mm of diameter and smooth and grooved surfaces during 7 days when the discs were separated in experimental groups (15 discs each group), in accordance with the decontamination method: LBI group (decontamination with InGaAlP laser, 660 nm, 30 mW, 45 J/cm2 in continuous way during 30 s); PDT group; TBO group (application of TBO during 60 s); controlled positive group (C) without any kind of treatment and controlled negative group (NC), sterile. The discs had been implanted in rats subcutaneous connective tissue and, after 7, 28 and 84 days, had been gotten 5 biopsies from each group for histology processing. Smooth and grooved implants did not differ on fibrosis grade and severity of the inflammatory infiltrate, however the perimaterial reaction tissue area was bigger on grooved implants (2,6 ± 3,7 x 106 µm2) that on smooth implants (1,9 ± 2,6 x 106 µm2). The C group was the one that showed the smallest reaction tissue fibrosis grade (1), but only was significant statistical difference (p = 0,0230) between C and NC groups. The severity of inflammatory infiltrate only significant differed between the groups on 7 days, when the NC group showed the smallest score (2) in comparison to the others groups (3). In relation to the perimaterial reaction tissue area, only were differences between the groups on 7 days, when C group showed bigger area than the others (9,11 ± 2,10 x 106 µm2), nevertheless, without significant statistical difference in Comparison to LBI and TBO groups. The area presented by PDT group (4,34 ± 1,49 x 106 µm2) only was inferior to NC group (0,65 ± 0,23 x 106 µm2). The reaction tissue thickness did not confirmed the difference between smooth and grooved surfaces discs (p = 0,2801), however it confirmed all the comparisons related to the area parameter, where NC group showed smaller thickness than the others only on 7 days. NC group showed smaller thickness than the others, but this difference was statistical significant only on 7 days. The other groups did not differ among them in this parameter. It was concluded that grooved and smooth implants contaminated by a microbian biofilm ant treated with LBI, PDT and TBO did not differed among them in relation to tissue inflammatory parameters, seems that is a light superiority of PDT in relation to the others only in the earliest period of observation. Nevertheless, with the time, all the treatments were equally to sterile implants in relation to perimaterial reaction tissue.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
SamiraSalmeron.pdf (9.36 Mbytes)
Fecha de Publicación
2011-08-22
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.