Titre en portugais
Avaliação in vitro do colar de implantes cone Morse de diferentes diâmetros sob cargas cêntrica e excêntrica: estudo por meio da Interferometria Eletrônica por Padrões de Speckle
Mots-clés en portugais
Cone Morse
Estresse mecânico
Força compressiva
Holografia
Implantes dentários
Interferometria
Resumé en portugais
Cargas oclusais excessivas podem promover tensões exageradas às estruturas de implantes/pilares/próteses, resultando em deformação, desadaptação ou mesmo fratura dos componentes e do próprio implante. Além disso, tais sobrecargas podem gerar necrose por compressão e perda óssea peri-implantar e, ainda, a ocorrência de microdeslocamentos na interface de conexão implante-pilar (I-P) ocasionando a formação de fendas propícias à colonização bacteriana, o que pode levar ao desenvolvimento de peri-implantite, contribuindo para o insucesso do conjunto. O presente estudo objetivou avaliar, por meio de uma técnica já amplamente utilizada na Engenharia e na Física a Interferometria Eletrônica por Padrões de Speckle (ESPI) as deformações a nível do colar de implantes com conexão cônica interna (cone Morse) de diferentes diâmetros (Kopp 4,3/5,5 mm e Neodent 3,5/5,0 mm), sob dois tipos (cêntrica e excêntrica) e intensidades (50 N e 100 N) de cargas, similares as quais os implantes ficam expostos durante a função mastigatória. Testou-se a hipótese de que a deformação do colar do implante seria inversamente proporcional ao diâmetro do implante. Além disso, discutiu-se vantagens e limitações da técnica ESPI frente a outros métodos comumente utilizados na avaliação de tensões. As maiores deformações foram sempre associadas às cargas excêntricas, com exceção do G1-Kopp. A maior deformação ocorreu no G1-Neodent sob carga excêntrica de 100 N e a menor, no G2-Neodent sob carga cêntrica de 50 N. A técnica ESPI mostrou-se altamente sensível e reprodutível para avaliar deformações a nivel do colar de implantes sendo possível confirmar a hipótese inicial, visto que a maior deformação esteve sempre associada aos implantes de menor diâmetro, em ambos os sistemas de implante pesquisados.
Titre en anglais
In vitro evaluation of the collar of Morse taper dental implants with different diameters under centric and eccentric loads: an electronic speckle pattern interferometry study
Mots-clés en anglais
Compressive load
Dental implants
Holography
Interferometry
Morse taper
Stress mechanical
Resumé en anglais
Occlusal overloading may promote excessive stresses to the implants/abutments/prosthesis, resulting in deformation, misfit or fracture of the components, until the implants. Futhermore, strains of the implant-abutment connection can cause the formation of microgaps propitious to bacterial colonization, which leads to the development of peri-implant sites, leading to the failure of the rehabilitation. This study aimed to evaluate, by a technique widely used in the engineering and physics Electronic Speckle Pattern Interferometry (ESPI) the strains related to the implant collar with internal tappered connection (Morse taper) with different diameters (Kopp 4.3/5.5 mm and Neodent 3,5/5,0 mm), under two types (centric and eccentric) and intensities (50 N and 100 N) of loads, similar which implants are exposed during masticatory function. We tested the hypothesis that the strains of the implants collar would be inversely proportional to the diameter of the implant. In addition, the survey discussed the advantages and limitations of the ESPI technique over other methods commonly used to assess strains. The largest strain was always associated with eccentric loads, except for G1-Kopp. The largest strain occurred in the G1-Neodent under the eccentric loads of 100 N and the lowest in the G2-Neodent under the centric loads of 50 N. The ESPI technique was a highly sensitive and reproducible to evaluate the strains at the level of the collar of implants being possible to confirm the initial hypothesis, since most strains was always associated with narrower-diameter implants in both implant systems studied.
 
AVERTISSEMENT - Regarde ce document est soumise à votre acceptation des conditions d'utilisation suivantes:
Ce document est uniquement à des fins privées pour la recherche et l'enseignement. Reproduction à des fins commerciales est interdite. Cette droits couvrent l'ensemble des données sur ce document ainsi que son contenu. Toute utilisation ou de copie de ce document, en totalité ou en partie, doit inclure le nom de l'auteur.
Date de Publication
2012-09-26
 
AVERTISSEMENT: Apprenez ce que sont des œvres dérivées cliquant ici.
Tous droits de la thèse/dissertation appartiennent aux auteurs
CeTI-SC/STI
Bibliothèque Numérique de Thèses et Mémoires de l'USP. Copyright © 2001-2024. Tous droits réservés.