Disertación de Maestría
DOI
https://doi.org/10.11606/D.23.2011.tde-30112011-163623
Documento
Autor
Nombre completo
Sergio Rodrigues Sizo
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2011
Director
Tribunal
Mori, Matsuyoshi (Presidente)
Contin, Ivo
Muramatsu, Mikiya
Título en portugués
Avaliação in vitro do colar de implantes cone Morse de diferentes diâmetros sob cargas cêntrica e excêntrica: estudo por meio da Interferometria Eletrônica por Padrões de Speckle
Palabras clave en portugués
Cone Morse
Estresse mecânico
Força compressiva
Holografia
Implantes dentários
Interferometria
Resumen en portugués
Cargas oclusais excessivas podem promover tensões exageradas às estruturas de implantes/pilares/próteses, resultando em deformação, desadaptação ou mesmo fratura dos componentes e do próprio implante. Além disso, tais sobrecargas podem gerar necrose por compressão e perda óssea peri-implantar e, ainda, a ocorrência de microdeslocamentos na interface de conexão implante-pilar (I-P) ocasionando a formação de fendas propícias à colonização bacteriana, o que pode levar ao desenvolvimento de peri-implantite, contribuindo para o insucesso do conjunto. O presente estudo objetivou avaliar, por meio de uma técnica já amplamente utilizada na Engenharia e na Física a Interferometria Eletrônica por Padrões de Speckle (ESPI) as deformações a nível do colar de implantes com conexão cônica interna (cone Morse) de diferentes diâmetros (Kopp 4,3/5,5 mm e Neodent 3,5/5,0 mm), sob dois tipos (cêntrica e excêntrica) e intensidades (50 N e 100 N) de cargas, similares as quais os implantes ficam expostos durante a função mastigatória. Testou-se a hipótese de que a deformação do colar do implante seria inversamente proporcional ao diâmetro do implante. Além disso, discutiu-se vantagens e limitações da técnica ESPI frente a outros métodos comumente utilizados na avaliação de tensões. As maiores deformações foram sempre associadas às cargas excêntricas, com exceção do G1-Kopp. A maior deformação ocorreu no G1-Neodent sob carga excêntrica de 100 N e a menor, no G2-Neodent sob carga cêntrica de 50 N. A técnica ESPI mostrou-se altamente sensível e reprodutível para avaliar deformações a nivel do colar de implantes sendo possível confirmar a hipótese inicial, visto que a maior deformação esteve sempre associada aos implantes de menor diâmetro, em ambos os sistemas de implante pesquisados.
Título en inglés
In vitro evaluation of the collar of Morse taper dental implants with different diameters under centric and eccentric loads: an electronic speckle pattern interferometry study
Palabras clave en inglés
Compressive load
Dental implants
Holography
Interferometry
Morse taper
Stress mechanical
Resumen en inglés
Occlusal overloading may promote excessive stresses to the implants/abutments/prosthesis, resulting in deformation, misfit or fracture of the components, until the implants. Futhermore, strains of the implant-abutment connection can cause the formation of microgaps propitious to bacterial colonization, which leads to the development of peri-implant sites, leading to the failure of the rehabilitation. This study aimed to evaluate, by a technique widely used in the engineering and physics Electronic Speckle Pattern Interferometry (ESPI) the strains related to the implant collar with internal tappered connection (Morse taper) with different diameters (Kopp 4.3/5.5 mm and Neodent 3,5/5,0 mm), under two types (centric and eccentric) and intensities (50 N and 100 N) of loads, similar which implants are exposed during masticatory function. We tested the hypothesis that the strains of the implants collar would be inversely proportional to the diameter of the implant. In addition, the survey discussed the advantages and limitations of the ESPI technique over other methods commonly used to assess strains. The largest strain was always associated with eccentric loads, except for G1-Kopp. The largest strain occurred in the G1-Neodent under the eccentric loads of 100 N and the lowest in the G2-Neodent under the centric loads of 50 N. The ESPI technique was a highly sensitive and reproducible to evaluate the strains at the level of the collar of implants being possible to confirm the initial hypothesis, since most strains was always associated with narrower-diameter implants in both implant systems studied.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2012-09-26
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.