Tesis Doctoral
DOI
https://doi.org/10.11606/T.23.2013.tde-19122013-195214
Documento
Autor
Nombre completo
Rodrigo Casasanta França
Dirección Electrónica
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Paulo, 2013
Director
Tribunal
Santos, Marcelo dos (Presidente)
Brzozowski, Margarete Elisabete
Chavez, Victor Elias Arana
Prokopowitsch, Igor
Salomão, José Antonio Saadi
Título en portugués
Avaliação do desgaste das lâminas de corte do sistema Reciproc através da utilização do microscópio eletrônico de varredura MEV
Palabras clave en portugués
Microscopia eletrônica de varredura
Sistema Rotatório
Resumen en portugués
Quando se utilizam os instrumentos rotatórios, oscilatórios e os recíprocos, de NiTi (Níquel-Titânio), cortando ou raspando as paredes do conduto radicular, estes também sofrem desgastes em suas lâminas de corte. Tal ocorrência determina atritos cada vez maiores, proporcionais à perda do corte de suas lâminas, diminuindo a eficiência do instrumento, exigindo mais torques e força de penetração, aumentando ainda mais o atrito e o esforço que este sofre para desgastar a dentina, predispondo acidentes. O presente estudo avaliou o desgaste nas lâminas de corte das limas Reciproc, após 3,6 e 9 usos em canais de dentes naturais. Foram utilizados 180 dentes pré-molares inferiores e 40 instrumentos R40. Após o preparo dos canais os instrumentos foram incluídos em corpos de prova e submetidos a cortes transversais a 3 mm e 6 mm, contados a partir da ponta do instrumento. Nas imagens dessas secções transversais obtidas no MEV foi mesurada a distância entre as extremidades das lâminas de corte. O desgaste dos instrumentos foi estabelecido pela diferença entre as medidas das distâncias dos instrumentos sem uso com a média de cada instrumento, nos diferentes grupos experimentais. Observou-se diferença significante, quando comparados os grupos de 6 e 9 usos com o grupo controle. Já o grupo com 3 utilizações não apresentou diferença estatisticamente significante em relação ao grupo controle. As diferenças percentuais de desgaste a 3 mm após 3, 6 e 9 utilizações foram de 5,96%, 8,01% e 18,09%, respectivamente, em relação ao Grupo controle. As diferenças percentuais de desgaste a 6 mm, após 3, 6 e 9 utilizações foram de 2,32%, 9,54% e 21,00%, respectivamente, em relação ao Grupo controle. Conclui-se, assim, que ocorre desgaste significante das lâminas de corte dos instrumentos Reciproc, já a partir do terceiro uso reforçando a recomendação do fabricante em utilizar cada instrumento em molar e, então, descartá-lo. Palavras-chave: Microscopia eletrônica de varredura. Sistema Rotatório.
Título en inglés
Evaluation of wear of the cutting blades Reciproc system using the scanning electron microscope SEM
Palabras clave en inglés
Reciproc System
Scanning electron microscopy
Resumen en inglés
The fracture of nickel-titanium instrument, whose consequences can range from the difficulty of its removal until the absolute inaccessibility to the apical portion of the canal, remains a situation to be avoided in the safest way possible. When using the rotary instruments, and reciprocal oscillatory, NiTi, cutting or scraping the walls of the root canal, they also suffer wear in its cutting blades. Such an occurrence determines friction increasing in proportion to the loss of cutting blades, reducing the efficiency of the instrument, requiring higher torque and penetration force, further increasing friction and effort that suffers when wearing the dentin. Different rotational speeds to the long axis of the active part of the instrument are thus provoked in areas of high friction, favoring fractures, regardless of the type of movement, even if it is rotating or reciprocating. The present study sought to assess whether there is wear on cutting blades System Reciproc, since this system works through reciprocal movement within the root canal. For that, we will instrumentation in selected teeth with the kinematic system recommended by the manufacturer at zero, three, six and nine uses, corresponding to a molar tooth, two molar teeth and three molar teeth after 10 uses in each experimental group (zero , three, six and nine) according to the number of uses. The instruments are inserted into a quadrangular centralized device in the same position, and fixed in resin, which have been cut into their cross initial 3 mm and 6 mm initial these selected samples with the cross section of the system is brought to the SEM, and thus wear measured by measuring the radius of your cutting blades, and thus statistically analyzed. The results showed statistically significant differences after successive instrumentations.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Fecha de Publicación
2014-02-11
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.