Master's Dissertation
DOI
https://doi.org/10.11606/D.23.2018.tde-05112018-123430
Document
Author
Full name
Jun Ho Kim
E-mail
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2018
Supervisor
Committee
Arita, Emiko Saito (President)
Côrtes, Arthur Rodriguez Gonzalez
Prado, Marina Cleia Palo
Raitz, Ricardo
Title in Portuguese
Comparação da precisão e acurácia de medidas lineares com diferentes protocolos de tomografia computadorizada por feixe cônico
Keywords in Portuguese
Artefatos
Defeitos ósseos periimplantares
Tomografia computadorizada por feixe cônico
Abstract in Portuguese
Uma das complicações mais importantes relacionadas com o tratamento de implante dentário é a peri-implantite. Devido a limitação pela sobreposição de estruturas da técnica periapical, nos casos da perda óssea peri-implantar nas faces vestibulares e linguais ou palatais necessita-se de um exame capaz de prover imagens em 3 dimensões para o correto diagnóstico, sendo as imagens de tomografia computadorizada por feixe cônico (TCFC) as mais indicadas. O objetivo deste estudo foi avaliar a acurácia, especificidade, sensibilidade, concordância inter e intraobservadores e determinar a influência dos artefatos na detecção dos defeitos ósseos peri-implantares simulados em mandíbulas humanas edêndulas secas. Foram utilizados dois aparelhos de TCFC com diferentes protocolos de aquisição: i-CAT Gendex CB-500(120kVp, 5mA) e ORTHOPANTOMOGRAPH OP300(90kVp, 6.3mA). Foram instalados 24 implantes dentários cilíndricos e executados dezesseis defeitos somente nas faces vestibulares, com os comprimentos que variaram de 0,5mm a 15mm. Somente cortes sagitais foram utilizados para avaliação qualitativa e quantitativa dos defeitos foi realizada por meio de Image J. Os resultados do teste de concordância intra e inter-observadores tanto nas avaliações qualitativas e quantitativas foram predominantemente fracos. Os valores de Área Sob a Curva (ASC) variaram de 0.51 a 0.72 no i-CAT Gendex CB-500; e no ORTHOPANTOMOGRAPH OP300 variaram de 0.54 a 0.75, demonstrando desempenho diagnóstico ligeiramente superior. Nas imagens obtidas em aparelho com 120kVp (i-CAT Gendex CB-500) foram observados maior contraste em relação ao aparelho 90kVp (ORTHOPANTOMOGRAPH® OP300); em aparelho com o menor kVp (OP300) formaram mais artefatos e estes, degradaram a qualidade das imagens e influenciaram o desempenho diagnóstico das TCFCs na detecção e nas mensurações de perda óssea peri-implantar. Os protocolos de aquisições diferentes podem influenciar no aumento ou redução da formação dos artefatos.
Title in English
Comparison of accuracy and precision of linear measurements with different cone beam computed tomographic protocols
Keywords in English
Artifacts
Cone beam computed tomographic
Peri-implant bone defects
Abstract in English
One of the most important complications related to dental implant treatment is periimplantitis. Due to the structures overlapping inherited to periapical technique, in cases of buccal and lingual or palatal bone loss in peri-implantitis, the 3 dimensions observation exam like cone beam computed tomography(CBCT) is most indicated. The objective of this study was to evaluate the accuracy, specificity, sensitivity, interand intra-observer agreement and to determine the influence of the artifacts in the detection of the peri-implant bone defects simulated in human dry mandibles. The two CBCT devices with different acquisition protocols were used in this study: i-CAT Gendex CB-500 (120kVp, 5mA) and ORTHOPANTOMOGRAPH OP300 (90kVp, 6.3mA). Twenty-four cylindrical dental implants were placed and 16 defects were prepared only on buccal faces, the defect lengths ranged from 0.5mm to 15mm. For qualitative and quantitative assessment, only sagittal slices were evaluated by the observers, using Image J software. The results of intra and inter-observer agreement tests in both assessments were generally poor. The AUC values ranged from 0.51 to 0.72 on the i-CAT Gendex CB-500; and in the ORTHOPANTOMOGRAPH OP300 ranged from 0.54 to 0.75, showing slightly higher diagnostic performance. Higher contrast was observed in the 120 kVp device's images (i-CAT Gendex CB-500) when compared with the 90 kVp device's image (ORTHOPANTOMOGRAPH® OP300). The low kVp device (OP300), formed more artifacts which degraded the image quality and influenced the CBCT diagnostic performance in the detection and measurement of peri-implant bone loss. Different CBCT acquisition protocols may influence increasing or reducing artefact formation.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.