Disertación de Maestría
DOI
https://doi.org/10.11606/D.18.2014.tde-19082014-090655
Documento
Autor
Nombre completo
Tales Roberto de Souza Santini
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Carlos, 2014
Director
Tribunal
Maciel, Carlos Dias (Presidente)
Carvalho, Aparecido Augusto de
Paterno, Aleksander Sade
Título en portugués
Projeto e análise de aplicações de circuladores ativos para a operação em frequências de ultrassom Doppler de ondas contínuas
Palabras clave en portugués
Análise do pior caso
CFOA
Circulador ativo
Faixa dinâmica
Largura de banda
Método de Monte-Carlo
PCB
Ultrassom Doppler
Resumen en portugués
Os circuladores tradicionais são amplamente utilizados em telecomunicações e defesa militar para o simultâneo envio e recepção de sinais por um único meio. Esses circuitos passivos, fabricados a partir de materiais ferromagnéticos, possuem a desvantagem do aumento de dimensões, peso e custos de fabricação com a diminuição da frequência de operação definida no projeto destes dispositivos, inviabilizando sua aplicação em frequências abaixo de 500 MHz. O circulador ativo surgiu como uma alternativa aos tradicionais, tendo aplicações em frequências desde o nível DC até a ordem de dezenas de gigahertz. As suas maiores aplicações ocorrem quando são necessários dispositivos compactos, de baixo custo e de baixa potência. Os primeiros circuitos propostos possuíam uma grande limitação em termos de frequência de operação e de potência entregue à carga. Entretanto, com os avanços tecnológicos na eletrônica, tais problemas podem ser amenizados atualmente. Neste trabalho é apresentado o desenvolvimento de um circuito circulador ativo para a utilização em instrumentação eletrônica, em particular para a operação em frequências na ordem das utilizadas em equipamentos de ultrassom Doppler de ondas contínuas, na faixa de 2 MHz a 10 MHz. As possíveis vantagens da implementação de circuladores em sistemas de ultrassom estão relacionadas ao incremento da relação sinal-ruído, aumento da área de recepção do transdutor, simplificação da construção do transdutor, simplificação do circuito de demodulação/ processamento, e maior isolação entre os circuitos de transmissão e recepção de sinais. Na fase inicial, o circulador ativo proposto é modelado por equacionamento, utilizando-se tanto o modelo ideal dos amplificadores operacionais como o seu modelo de resposta em frequência. Simulações computacionais foram executadas para confirmar a validade do equacionamento. Um circuito montado em placa de prototipagem rápida foi apresentado, e testes de prova de conceito em baixas frequências foram realizados, mostrando uma grande semelhança entre o teórico, o simulado e o experimental. A segunda parte contou com o projeto do circuito circulador para a operação em maiores frequências. O circuito proposto é composto por três amplificadores operacionais de realimentação por corrente e vários componentes passivos. Uma análise de sensibilidade utilizando os métodos de Monte-Carlo e análise do pior caso foi aplicada, resultando em um perfil de comportamento frente às variações dos componentes do circuito e às variações da impedância de carga. Uma placa de circuito impressa foi projetada, utilizando-se de boas práticas de leiaute para a operação em altas frequências. Neste circuito montado, foram realizados os seguintes testes e medições: comportamento no domínio do tempo, faixa dinâmica, nível de isolação em relação à amplitude do sinal, largura de banda, levantamento dos parâmetros de espalhamento, e envio e recepção de sinais por transdutor de ultrassom Doppler de ondas contínuas. Os resultados dos testes de desempenho foram satisfatórios, apresentando uma banda de transmissão de sinais para frequências de 100 MHz, isolação entre portas não consecutivas de 39 dB na frequência de interesse para ultrassom Doppler e isolação maior que 20 dB para frequências de até 35 MHz. A faixa dinâmica excedeu a tensão de 5 Vpp, e o circuito teve bom comportamento no envio e na recepção simultânea de sinais pelo transdutor de ultrassom.
Título en inglés
Design and application analysis of active circulators for operation in frequencies of continuous-wave Doppler ultrasound
Palabras clave en inglés
Active circulator
Bandwidth
CFOA
Doppler ultrasound
Dynamic range
Monte-Carlo method
PCB
Worst-case analysis
Resumen en inglés
Traditional circulators are widely used in both telecommunications and military defense for sending and receiving signals simultaneously through a single medium. These passive circuits which are manufactured from ferromagnetic materials, have the disadvantages of having suffered an increase in dimensions, weight, and manufacturing costs along with the decrease in the operation frequency established in the designs of such devices, thus preventing their useful employment in frequencies below 500 MHz. The active circulator emerged as an alternative to the traditional ones, and has applications on frequencies ranging from a DC level to levels involving dozens of gigahertz. It is applicable when compact devices are made necessary, at a low cost, and for low frequencies. The first circuits to be introduced had a major limitation in terms of operating frequency and power delivered to the load. However, due to technological advances in electronics, problems such as the aforementioned can now be minimized. This research work presents the development of an active circulator circuit to be used in electronic instrumentation, particularly for operation at frequencies such as those used in continuous wave Doppler ultrasound equipment, ranging from 2 MHz to 10 MHz. The advantages made possible by implementing ultrasound systems with circulators are related to an increase in the signal-to-noise ratio, an increase in the transducers reception area, a simplified construction of the transducer, simplification of the demodulation/processing circuit, and a greater isolation between the transmission circuits and signal reception. In the initial phase, the proposed active circulator was modeled by means of an equating method, using both the ideal model of operational amplifiers and the model of frequency response. Computer simulations were carried out in order to confirm the validity of the equating method. A circuit mounted upon a breadboard was introduced and proof of concept assessments were performed at low frequencies, showing a great similarity among the theoretical, simulated and experimented data. The second phase is when the circulator circuits design was developed in order make its operation at higher frequencies possible. The proposed circuit is comprised of three currentfeedback operational amplifiers and several passive components. A sensitivity analysis was carried out using Monte-Carlo methods and worst-case analyses, resulting in a certain behavioral profile influenced by variations in circuit components and variations in load impedance. A printed circuit board was designed, employing good practice layout standards so that operation at high frequencies would be achieved. The following evaluations and measurements were performed on the circuit that was assembled: time domain behavior, dynamic range, isolation level relative to signal amplitude, bandwidth, survey of the scattering parameters, and transmission and reception of signals by a continuous wave Doppler ultrasound transducer. The results of the performance tests were satisfactory, presenting a 100 MHz signal transmission band, isolation between non-consecutive ports of 39 dB at the frequency of interest to the Doppler ultrasound, and an isolation greater than 20 dB for frequencies of up to 35 MHz. The dynamic range exceeded the 5Vpp and the circuit performed satisfactorily in the simultaneous transmission and reception of signals through the ultrasound's transducer.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Tales.pdf (8.99 Mbytes)
Fecha de Publicación
2014-08-28
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.