• JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
  • JoomlaWorks Simple Image Rotator
 
  Bookmark and Share
 
 
Disertación de Maestría
DOI
https://doi.org/10.11606/D.18.2015.tde-24082015-074834
Documento
Autor
Nombre completo
Lucas Rodrigues Borges
Instituto/Escuela/Facultad
Área de Conocimiento
Fecha de Defensa
Publicación
São Carlos, 2015
Director
Tribunal
Vieira, Marcelo Andrade da Costa (Presidente)
Bakic, Predrag
Mascarenhas, Nelson Delfino d'Ávila
Título en portugués
Método de simulação da redução da dose de radiação na mamografia digital a partir da análise das características do ruído dos equipamentos mamográficos
Palabras clave en portugués
Mamografia digital
Redução de dose de radiação
Ruído quântico
Transformada de Anscombe
Resumen en portugués
Este trabalho tem como objetivo o desenvolvimento de um novo método para a simulação de redução da dose de radiação em imagens mamográficas clínicas. Assim, estudos sobre a influência da redução da dose de radiação no diagnóstico do câncer de mama podem ser realizados sem que o paciente se exponha a doses extras de radiação. Uma análise preliminar foi realizada para a caracterização do ruído produzido pelo equipamento mamográfico no processo de aquisição da imagem. Essa análise evidenciou a importância de um método local de simulação, uma vez que o ruído depende da posição ao longo do campo. O novo método proposto consiste em ajustar os níveis de cinza e adicionar uma máscara de ruído Poisson, dependente do sinal, nas imagens clínicas adquiridas com a dose de radiação padrão, simulando sua aquisição com doses de radiação reduzidas. A dependência entre ruído e sinal foi criada com o uso da transformada de Anscombe. O desempenho do método proposto foi avaliado utilizando-se imagens mamográficas de um phantom antropomórfico obtidas com diferentes doses de radiação. As imagens simuladas pelo método proposto foram comparadas com as imagens reais. A similaridade entre os espectros de ruído permitiu a comparação de métricas locais da imagem. O erro percentual entre os níveis de cinza das imagens reais e simuladas se manteve inferior a 1%. O ruído adicionado manteve um erro percentual inferior a 1%. Testes de t-Student mostraram que não existe diferença estatística significante (p < 0,05) entre as imagens reais e simuladas pelo método proposto.
Título en inglés
Method for simulating dose reduction in digital mammography through the analysis of the noise characteristics of the mammographic equipment
Palabras clave en inglés
Anscombe transformation
Digital mammography
Quantum noise
Reduction of the radiation dose
Resumen en inglés
This work aims to develop a new method for simulating reduction of the radiation dose in clinical digital mammography. Using such method, studies regarding the influence of dose reduction in cancer diagnosis can be performed without unnecessary exposure of patients to X-ray radiation. A preliminary study characterized the noise produced by the digital mammography equipment during the acquisition process. This analysis emphasized the importance of simulating noise locally, since noise is dependent on the spatial position of the pixel. Therefore, the proposed method consists of adjusting the gray levels and adding signal-dependent Poisson noise to images acquired at the standard radiation dose. Dependency between noise and signal was created using the Anscombe transformation. The performance of the proposed method was evaluated using mammographic images of an anthropomorphic phantom acquired at different radiation doses. Images simulated using the proposed method were compared to real images acquired using the clinical equipment. Similarity between noise power spectra and local metrics validated the similarity between images. The gray level of the simulated and real images were compared using local mean and reported averaged errors smaller than 1%. The added noise was also compared and the averaged error was smaller than 1%. Statistical Student's t-test tests showed no statistical difference (p < 0.05) between real images and the ones simulated using the proposed method.
 
ADVERTENCIA - La consulta de este documento queda condicionada a la aceptación de las siguientes condiciones de uso:
Este documento es únicamente para usos privados enmarcados en actividades de investigación y docencia. No se autoriza su reproducción con finalidades de lucro. Esta reserva de derechos afecta tanto los datos del documento como a sus contenidos. En la utilización o cita de partes del documento es obligado indicar el nombre de la persona autora.
Lucas.pdf (7.05 Mbytes)
Fecha de Publicación
2015-08-27
 
ADVERTENCIA: Aprenda que son los trabajos derivados haciendo clic aquí.
Todos los derechos de la tesis/disertación pertenecen a los autores
CeTI-SC/STI
Biblioteca Digital de Tesis y Disertaciones de la USP. Copyright © 2001-2024. Todos los derechos reservados.