Título em português
Estudo e simulação de métodos de colimação para melhoria de nitidez da imagem mamográfica
Palavras-chave em português
Colimador
Imagem médica
Mamógrafo
Simulação de métodos
Resumo em português
Neste trabalho estamos propondo diminuir o tamanho do ponto focal usando um colimador externo ao aparelho, no lado do catodo, mascarando assim, parte do ponto focal. Esta localização permite utilizar a região onde há maior intensidade de radiação por efeito "heel". Este método tem também a vantagem de diminuir o efeito Compton. As localizações do colimador foram encontradas por simulação computacional de acordo com o seguinte compromisso: maior faixa do filme com ponto focal reduzido sem comprometer a intensidade do feixe de radiação. A simulação computacional foi programada em linguagem pascal (Delphi) considerando que o ponto focal possui forma e distribuição de intensidade variadas. Os parâmetros para a simulação foram utilizados para a realização de um dispositivo experimental. As imagens obtidas com este dispositivo foram comparadas às imagens simuladas afim de comprovar a validade do modelo.
Título em inglês
not available
Palavras-chave em inglês
not available
Resumo em inglês
Throughout this work we intend to decrease the focal spot size putting an extemal collimator on the side of the cathode beneath the mammography device, this concealing part of the focal spot. This setting permits the use of the area where there is more radiation intensity by "heel" effect. This method has the advantage of decreasing the Compton effect too. The setting of the collimator were determined by computer simulation in the accordance in the following commitment: a larger band of the film and reduced focal spot size. These settings were determined without committing the radiation beam intensity. The computer simulation was programmed using pascal (Delphi) taking in a count that the focal spot has varied intensity distribution and geometry. The parameters applied for the simulation were used for the accomplishment of an experimental device. The images obtained with this device were compared to the simulate images so as to prove the validity of the model.
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
Data de Publicação
2017-10-30
 
AVISO: Saiba o que são os trabalhos decorrentes clicando aqui.
Todos os direitos da tese/dissertação são de seus autores
CeTI-SC/STI
Biblioteca Digital de Teses e Dissertações da USP. Copyright © 2001-2024. Todos os direitos reservados.