Doctoral Thesis
DOI
https://doi.org/10.11606/T.18.2016.tde-19022016-151911
Document
Author
Full name
Ricardo Toshiyuki Irita
Institute/School/College
Knowledge Area
Date of Defense
Published
São Carlos, 2003
Supervisor
Committee
Slaets, Annie France Frère (President)
Furuie, Sérgio Shiguemi
Infantosi, Antonio Fernando Catelli
Moraes, José Carlos Teixeira de Barros
Silva Filho, Antonio Carlos Roque da
Title in Portuguese
Técnica de colimação para otimizar a aquisição e o processamento de imagens mamográficas
Keywords in Portuguese
Espalhamento
Modelagem
Quantificação da radiação X
Raios-X
Simulação da radiação X
Slit mamográfico modificado
Abstract in Portuguese
Para melhorar a visualização das pequenas estruturas anatômicas importantes para o diagnóstico do câncer de mama e otimizar o processamento das mamografias pelos sistemas computadorizados de auxílio ao diagnóstico (CAD), foi desenvolvido um dispositivo, baseado na tecnologia slit, que melhora a aquisição dos mamogramas. Este dispositivo reduz a radiação espalhada e o tamanho do foco e foi projetado a partir de um modelo computacional. O modelo adotado permite quantificar o valor desses parâmetros para qualquer sistema radiológico, qualquer espessura de tecidos moles radiografada e qualquer tensão aplicada ao tubo de raios-X. O dispositivo foi implementado e testado, quantificando as melhorias obtidas. As imagens geradas foram comparadas com as fornecidas pelos sistemas mamográficos convencionais. O modelo serviu para estudar também a interferência do espalhamento sobre o desempenho dos algoritmos usados nos sistemas de diagnóstico auxiliado por computador (CAD).
Title in English
Collimation techniques for optimization of mammography image acquisition and processing
Keywords in English
Computer modeling
Modified mammography slit
Scatter radiation
X-ray
X-ray quantification
X-ray simulation
Abstract in English
In order to improve the visualization of the small and important anatomical structures for breast cancer diagnosis and to optimize the image processing of mammograms by the computer aided diagnosis systems (CAD), a device that comes from slit technology was developed to improve the acquisition of mammograms. This device reduces the amount of scattered radiation and the focus size and it can be projected from a computational model. This model allows quantifying the value of those parameters, scattered radiation and the focus size, for any radiological system, any X-rayed soft tissue thickness and any tension applied to the X-ray tube. The device was implemented and tested quantifying the obtained improvements. The generated images were compared to the supplied for conventional mammography systems. The model was also good for studying interference on the algorithms used in the computer aided diagnosis systems (CAD).
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2016-02-19
 
WARNING: Learn what derived works are clicking here.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.