Master's Dissertation
DOI
https://doi.org/10.11606/D.10.2012.tde-07062013-152847
Document
Author
Full name
Jaqueline França dos Santos
Institute/School/College
Knowledge Area
Date of Defense
Published
São Paulo, 2012
Supervisor
Committee
Ferrigno, Cássio Ricardo Auada (President)
Mascarenhas, Nilva Maria Freres
Stopiglia, Angelo João
Title in Portuguese
Avaliação mecânica e de composição de liga metálica de dois tamanhos de anéis para fixador externo circular de três diferentes fabricantes para cães
Keywords in Portuguese
Anel para fixador
Compressão axial
Ilizarov
TTPIXE
Abstract in Portuguese
Vários implantes cirúrgicos têm sido desenvolvidos para estabilizar fraturas em seres humanos e animais. O fixador esquelético externo circular de Ilizarov é método de fixação amplamente utilizado para tratamento de fraturas complexas, desvios angulares, bem como no alongamento ósseo. O anel destaca-se como componente fundamental desse dispositivo e as cargas de compressão no mesmo plano do anel são devidas principalmente a fios ligados de um lado ao outro do implante. O estudo teve como objetivo comparar a resistência à compressão axial de anéis de Ilizarov e analisar a composição da liga metálica desses anéis. Foram testados dois diâmetros de anéis de três fabricantes nacionais para os ensaios biomecânicos de compressão axial, divididos em dois grupos com quinze corpos de prova. Os testes foram realizados em máquina universal de ensaios Kratos® modelo KE 3.000MP, dotada de célula de carga de 3000N com velocidade de ensaio de 10 milímetros por minuto. Os parâmetros força e deformação foram enviados ao computador IBM®-PC através do sistema de aquisição analógica Lynx® modelo ADS-2000, que permitiu a gravação dos parâmetros para posterior análise. Também avaliou-se a composição elementar dos anéis através do método TTPIXE (partículas induzidas pela emissão de raios X em alvos espessos) em PIXE externo. Os espectros foram acumulados em buffer-multicanal conectado a um micro-computador pessoal tipo IBM-PC®. A redução de espectros de raios-X foi realizada por meio do programa AXIL. Concluiu-se que os anéis para fixador esquelético externo circular são compostos de liga metálica de alumínio e que existe diferença entre os fabricantes nacionais no ensaio mecânico de compressão axial, sendo que o fabricante A é mais rígido que os outros fabricantes para os dois diâmetros (60 e 80 milímetros) de anel e mais resistente que os outros fabricantes para o anel de 80 milímetros.
Title in English
Evaluation of mechanical and metal alloy composition of two sizes of circular external fixator rings for three different manufacturers for dogs
Keywords in English
Axial compression
Ilizarov
Ring fixator
TTPIXE
Abstract in English
Various surgical implants have been developed to stabilize fractures in humans and animals. Ilizarov's circular external skeletal fixator is a fixation method widely used for the treatment of complex fractures, angular deviation, as well as in bone lengthening. The ring stands out as a fundamental component of the device and compression loads in the same plane of the ring are due mainly to wires connected from one side to another of the implant. The study aimed to compare the resistance to axial compression on Ilizarovs rings and analyze the alloy composition of these rings. Two diameters were tested of three domestic ring manufacturers for biomechanical axial compression, divided into two groups with fifteen specimens each group. The tests were performed in a universal test machine Kratos® model 3.000MP KE, equipped with a 3000N load cell with a test speed of 10 mm per minute. The strength and deformation parameters were sent to the IBM®-PC computer through the Lynx® analog acquisition system model ADS-2000, which allowed the recording of the parameters for further analysis. Also, the rings alloy composition was evaluated using the TTPIXE method (thick-target particle induced X-ray emission) in external PIXE. The spectra were accumulated in a multi-channel buffer connected to an IBM®-PC type personal microcomputer. The reduction of X-ray spectra was performed using the AXIL program. In conclusion the rings was made from aluminum alloy, witch differs in composition among different fabricants, also the rings showed differences performance in mechanical testing of axial compression, and the manufacturer A is more rigid than the other manufacturers for the two diameters (60 and 80 mm) ring and tougher than the other manufacturers to the 80 mm ring.
 
WARNING - Viewing this document is conditioned on your acceptance of the following terms of use:
This document is only for private use for research and teaching activities. Reproduction for commercial use is forbidden. This rights cover the whole data about this document as well as its contents. Any uses or copies of this document in whole or in part must include the author's name.
Publishing Date
2013-09-25
 
WARNING: The material described below relates to works resulting from this thesis or dissertation. The contents of these works are the author's responsibility.
  • DAL-BO, I. S., et al. Infecção óssea após osteotomia para tratamento da ruptura de ligamento cruzado em cães. Acta Scientiae Veterinariae , 2013, vol. 41, p. 1148.
All rights of the thesis/dissertation are from the authors
CeTI-SC/STI
Digital Library of Theses and Dissertations of USP. Copyright © 2001-2024. All rights reserved.