Dissertação de Mestrado
DOI
https://doi.org/10.11606/D.10.2012.tde-07062013-152847
Documento
Autor
Nome completo
Jaqueline França dos Santos
Unidade da USP
Área do Conhecimento
Data de Defesa
Imprenta
São Paulo, 2012
Orientador
Banca examinadora
Ferrigno, Cássio Ricardo Auada (Presidente)
Mascarenhas, Nilva Maria Freres
Stopiglia, Angelo João
Título em português
Avaliação mecânica e de composição de liga metálica de dois tamanhos de anéis para fixador externo circular de três diferentes fabricantes para cães
Palavras-chave em português
Anel para fixador
Compressão axial
Ilizarov
TTPIXE
Resumo em português
Vários implantes cirúrgicos têm sido desenvolvidos para estabilizar fraturas em seres humanos e animais. O fixador esquelético externo circular de Ilizarov é método de fixação amplamente utilizado para tratamento de fraturas complexas, desvios angulares, bem como no alongamento ósseo. O anel destaca-se como componente fundamental desse dispositivo e as cargas de compressão no mesmo plano do anel são devidas principalmente a fios ligados de um lado ao outro do implante. O estudo teve como objetivo comparar a resistência à compressão axial de anéis de Ilizarov e analisar a composição da liga metálica desses anéis. Foram testados dois diâmetros de anéis de três fabricantes nacionais para os ensaios biomecânicos de compressão axial, divididos em dois grupos com quinze corpos de prova. Os testes foram realizados em máquina universal de ensaios Kratos® modelo KE 3.000MP, dotada de célula de carga de 3000N com velocidade de ensaio de 10 milímetros por minuto. Os parâmetros força e deformação foram enviados ao computador IBM®-PC através do sistema de aquisição analógica Lynx® modelo ADS-2000, que permitiu a gravação dos parâmetros para posterior análise. Também avaliou-se a composição elementar dos anéis através do método TTPIXE (partículas induzidas pela emissão de raios X em alvos espessos) em PIXE externo. Os espectros foram acumulados em buffer-multicanal conectado a um micro-computador pessoal tipo IBM-PC®. A redução de espectros de raios-X foi realizada por meio do programa AXIL. Concluiu-se que os anéis para fixador esquelético externo circular são compostos de liga metálica de alumínio e que existe diferença entre os fabricantes nacionais no ensaio mecânico de compressão axial, sendo que o fabricante A é mais rígido que os outros fabricantes para os dois diâmetros (60 e 80 milímetros) de anel e mais resistente que os outros fabricantes para o anel de 80 milímetros.
Título em inglês
Evaluation of mechanical and metal alloy composition of two sizes of circular external fixator rings for three different manufacturers for dogs
Palavras-chave em inglês
Axial compression
Ilizarov
Ring fixator
TTPIXE
Resumo em inglês
Various surgical implants have been developed to stabilize fractures in humans and animals. Ilizarov's circular external skeletal fixator is a fixation method widely used for the treatment of complex fractures, angular deviation, as well as in bone lengthening. The ring stands out as a fundamental component of the device and compression loads in the same plane of the ring are due mainly to wires connected from one side to another of the implant. The study aimed to compare the resistance to axial compression on Ilizarovs rings and analyze the alloy composition of these rings. Two diameters were tested of three domestic ring manufacturers for biomechanical axial compression, divided into two groups with fifteen specimens each group. The tests were performed in a universal test machine Kratos® model 3.000MP KE, equipped with a 3000N load cell with a test speed of 10 mm per minute. The strength and deformation parameters were sent to the IBM®-PC computer through the Lynx® analog acquisition system model ADS-2000, which allowed the recording of the parameters for further analysis. Also, the rings alloy composition was evaluated using the TTPIXE method (thick-target particle induced X-ray emission) in external PIXE. The spectra were accumulated in a multi-channel buffer connected to an IBM®-PC type personal microcomputer. The reduction of X-ray spectra was performed using the AXIL program. In conclusion the rings was made from aluminum alloy, witch differs in composition among different fabricants, also the rings showed differences performance in mechanical testing of axial compression, and the manufacturer A is more rigid than the other manufacturers for the two diameters (60 and 80 mm) ring and tougher than the other manufacturers to the 80 mm ring.
 
AVISO - A consulta a este documento fica condicionada na aceitação das seguintes condições de uso:
Este trabalho é somente para uso privado de atividades de pesquisa e ensino. Não é autorizada sua reprodução para quaisquer fins lucrativos. Esta reserva de direitos abrange a todos os dados do documento bem como seu conteúdo. Na utilização ou citação de partes do documento é obrigatório mencionar nome da pessoa autora do trabalho.
Data de Publicação
2013-09-25
 
AVISO: O material descrito abaixo refere-se a trabalhos decorrentes desta tese ou dissertação. O conteúdo desses trabalhos é de inteira responsabilidade do autor da tese ou dissertação.
  • DAL-BO, I. S., et al. Infecção óssea após osteotomia para tratamento da ruptura de ligamento cruzado em cães. Acta Scientiae Veterinariae , 2013, vol. 41, p. 1148.
Todos os direitos da tese/dissertação são de seus autores
CeTI-SC/STI
Biblioteca Digital de Teses e Dissertações da USP. Copyright © 2001-2024. Todos os direitos reservados.